background image

4.8.  Mass Air Flow (MAF) Sensor (fig 9) 

4.8.1. A mass air flow sensor converts air flow readings into one of the following, depending on the  

 

sensor:

 

a) a dC voltage

 

b) a low frequency signal

 

c) a high frequency signal

 

The TA200 is suitable for testing those sensors with a dC voltage or low frequency signal  

 

output. 

4.8.2. Connect the dC voltage type MAF sensor in the dC voltage testing fashion (see section 4.1)  

 

and set the rotary switch on the meter to 20V dC. For a low frequency signal type MAF 

 

sensor, connect the meter in the RMPx10 testing fashion (see section 4.7) and choose the  

 

appropriate setting for number of cylinders. 

4.8.3. Connect the black test lead probe to the ground terminal of the sensor and connect the red  

 

one as illustrated in fig 9.

4.8.4. Switch on the meter and the vehicles ignition but 

DO NOT

 start the engine. 

4.8.5. 

DC Voltage type sensor: 

The displayed value should usually be less than or equal to 1V  

 

(consult the vehicle manufacturer’s specifications).

4.8.6. 

Frequency type sensor: 

Refer to the vehicle manufacturer’s specifications to compare  

 

against your readings. 

4.9.  Other functions

4.9.1. The TA200 can also be used to test the following automotive components:

 

Oxygen sensors, fuel injectors, temperature sensors, position sensors, MAP and baro  

 

sensors.

4.9.2. Please refer to the vehicle manufacturer’s manual for specific test procedures. 

Fig 9

5. 

MAINTENANCE

  

WARNING! DO NOT attempt to repair or service your meter unless you are qualified to do so and have the relevant calibration,  

 

performance test and service information.  

5.1.

 

General maintenance:

5.1.1. Periodically wipe the case with a damp cloth and mild detergent. 

DO NOT

 use abrasives or solvents. 

5.1.2. Clean the terminals with a cotton bud and mild detergent to prevent the build up of dirt/moisture in the terminals which may affect readings. 

5.1.3. Turn the meter off when not in use and remove the battery if the meter is not to be used for some time. 

5.2.

 

To replace the fuses:

5.2.1. Turn the meter off and disconnect any leads/equipment from the terminals.

5.2.2. Remove the protective rubber jacket from the meter.

5.2.3. Remove the 3 larger screws from the rear of the case and open the meter. 

5.2.4. Remove the fuse in question and replace with a fuse of identical type and specification:

 

Fuse 1: Ce 315mA, 250V, fast type, 5x20mm

 

Fuse 2: Ce 10A, 250V, fast type, 5x20mm 

5.2.5. Close the meter case and replace the three screws and rubber jacket. 

5.3.

 

To replace the battery:

5.3.1. Turn the meter off and disconnect and leads/equipment from the terminals. 

5.3.2. Remove the small screw holding the battery compartment lid, then remove the lid and the old battery.

5.3.3. Place a new 9V (PP3) battery into the meter taking care to observe the correct polarity.  

5.3.4. Replace the battery compartment lid and the holding screw.  

6. 

TECHNICAL SPECIFICATIONS

Range Resolution Accuracy Overload protection

200mV

0.1mV

2V

1mV

20V

10mV

200V

100mV

230V AC

±(0.5%+5)

 1000V dC or 

 750V AC continuous

1000V

1V

±(0.8%+5)

dC Voltage

Range Resolution Accuracy Overload protection

2V

1mV

20V

10mV

200V

100mV

±(0.8%+5)

1000V dC or 

750V AC continuous

750V

1V

±(1.0%+4)

AC Voltage

Range Resolution Accuracy Overload protection

200

0.1

2k

1

20k

10

200k

100

600Vp

±(0.8%+5)

2M

1k

Resistance

2M

10k

±(1.5%+5)

Range Resolution Overload protection

1mV

600Vp

Continuity testing

Input impedance: 10M

Input impedance: 10M

Frequency response: 40Hz ~ 400Hz

 TA200      Issue No: 3 - 19/12/11

Original Language Version

Содержание TA200

Страница 1: ... meter terminals When the scale of the value to be measured is unknown set the selector to the highest range available Before rotating the rotary switch to change functions disconnect test leads from the circuit under test WARNING Never perform resistance measurements on live circuits Always be careful when working with voltages above 60Vdc or 30Vac rms Keep your fingers behind the probe guards wh...

Страница 2: ...ates an AC voltage or current The displayed value is the mean value Indicates a negative reading Diode test Continuity buzzer is on Data hold function is active Indicates the requirement for test leads to be connected Ω Ohm The unit of resistance kΩ kilohm 1x103 or 1000 ohms MΩ Megaohm 1x106 or 1 000 000 ohms V Volts The unit of voltage mV Millivolt 1x10 3 or 0 001 volts A Amperes amps The unit of...

Страница 3: ...nt path to be tested connect the red test lead to the positive side of the break and the black test lead to the negative side of the break i e connect the meter in series 4 2 6 Turn on reconnect power to the circuit The measured value is shown on the display When the measurement has been completed disconnect the test leads from the circuit being tested Note Within the 10A range the meter must not ...

Страница 4: ... tested is open Open circuit voltage is approx 2 7V When continuity testing has been completed disconnect the connection between the testing leads and the circuit under test 4 6 Dwell testing fig 7 4 6 1 This function can be used to test the dwell of the cut off switch points or electronic of an ignition system 4 6 1 1 Turn the rotary switch to the appropriate setting within the DWELL function are...

Страница 5: ...cloth and mild detergent DO NOT use abrasives or solvents 5 1 2 Clean the terminals with a cotton bud and mild detergent to prevent the build up of dirt moisture in the terminals which may affect readings 5 1 3 Turn the meter off when not in use and remove the battery if the meter is not to be used for some time 5 2 To replace the fuses 5 2 1 Turn the meter off and disconnect any leads equipment f...

Страница 6: ... 5 Tach rotation speed testing Input amplitude 10V in direct impulse 0 5mS in width Input amplitude 10V in direct impulse 0 5mS in width Maximum tach 10000 RPM displayed as RPM x10 Environmental Protection Recycle unwanted materials instead of disposing of them as waste All tools accessories and packaging should be sorted taken to a recycle centre and disposed of in a manner which is compatible wi...

Отзывы: