5.3.   Accurate Measurement

5.3.1. Follow procedure 5.1.1 through 5.1.4.

5.3.2. The bore or gap machined finish must be clean with surface texture 3.2µm to 6.4µm or better than for fiducial indication.

5.3.3. offer the anvil end into the bore slightly tilted (fig.8) with the sprung loaded carriage entering just ahead of the anvil tip. rock the dial    

 

gauge tube to the upright position (fig.9) and beyond, observing the pointer sweep. The three point location offered by the two skids on  

 

the spring loaded carriage and the anvil tip will centralise the head.

5.3.4. rotate the bezel such that the "0" [zero] aligns with the largest sweep position of the pointer (fig.10).It is essential to observe the total    

 

sweep of the dial pointer using the small dial. remember from the initial setting the pointer had rotated one full cycle. Mark the position  

 

on the component of where the measurement was taken, if required.

5.3.5. With a micrometer, measure across the anvil and the nib (fig.11), rotate the barrel of the micrometer until the dial pointer aligns with the  

 

same "0" [zero] as the measured bore. Take the reading from the micrometer and record.

5.3.6. With the bore dial gauge now set, the bore can now be measured for taper and ovality by reference back to the initial datum "0" [zero]   

 

reading and counting the 0.01mm divisions between datum and new. note! the larger the deflection the smaller the bore. 

fig.8

fig.10

fig.9

fig.9

fig.11

"offer the anvil end into the bore slightly tilted...."

"rock the dial gauge tube to the upright position....."

"rotate the bezel such that the

"0" [zero] aligns....."

"With a micrometer, measure across the anvil and the nib......"

6. MAINTENANCE

6.1.

  Keep all components dry and clean with a soft micro fibre cloth.

6.2. 

return all items to the presentation case after use.

6.3. 

store indoors in a temperature controlled dry environment, circa 21°C.

6.4. 

This is a precision instrument intended for use by engineers and engineering inspectors, keep out of reach of children.

Component

Component

  dBG509     Issue: 1 - 21/06/15

Original Language Version

 © Jack sealey limited

A method of support when using a micrometer

Содержание DBG509

Страница 1: ...R THE PURPOSE FOR WHICH IT IS INTENDED FAILURE TO DO SO MAY CAUSE DAMAGE AND OR PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE Thank you for purchasing a Sealey product Manufactured to a high standard this product will if used according to these instructions and properly maintained give you years of trouble free performance 2 introduction Precision mec...

Страница 2: ... anvil 44 plus spacer ring 2 from item 1 in fig 5 Assemble as shown above in fig 1 5 2 4 Maneouvre the gauge anvil and nib inside the measuring gap of the micrometer fig 11 This activity could be eased by assistance or carefully clamping the probe in a retort stand or similar 5 2 5 Observe the large dial pointer movement It is essential that the axis of the gauge anvil and spring loaded nib are ce...

Страница 3: ... anvil and the nib fig 11 rotate the barrel of the micrometer until the dial pointer aligns with the same 0 zero as the measured bore Take the reading from the micrometer and record 5 3 6 With the bore dial gauge now set the bore can now be measured for taper and ovality by reference back to the initial datum 0 zero reading and counting the 0 01mm divisions between datum and new Note the larger th...

Страница 4: ...e is 12 months from purchase date proof of which will be required for any claim 01284 757500 01284 703534 sales sealey co uk Sole UK Distributor Sealey Group Kempson Way Suffolk Business Park Bury St Edmunds Suffolk IP32 7AR www sealey co uk x Anvil size Spacer ring 34 5 35 5 35 No 35 5 36 5 35 1x1mm 36 5 37 5 35 1x2mm 37 5 38 5 38 No 38 5 39 5 38 1x1mm 39 5 40 5 38 1x2mm 40 5 41 5 41 No 41 5 42 5...

Отзывы: