5. OPERATION

5.1.   Setting up (The dial ball gauge is a comparator not a measuring tool)

5.1.1. Measure the bore diameter or gap to be measured to the nearest millimetre with a rule or vernier callipers. The bore or gap    

 

 

must range between 35mm and 50mm nominally with this tool. (up to 52.5 with spacers)

5.1.2. select the range of parts required, anvils and spacers using the table on the last page as a guide. 

5.1.3. Assemble the dial bore gauge with selected parts as shown in fig.1. Ensure all components are clean when assembled, use the  

 

 

supplied spanner. 

DO NOT

 overtighten

5.1.4. Insert the dial stylus into the probe body as shown in fig.4. The stylus will meet with resistance internally and the dial pointer will  

 

 

begin to rotate clockwise. one complete cycle of the pointer is recommended for registration and will be indicated by the secondary dial  

 

indicator annotated 1-5. The small pointer should now be indicating "1" [one].

5.1.5. Clamp the dial stem with the thumb screw. 

DO NOT

 overtighten. 

 

5.2.   Calibration 

5.2.1. depending upon accuracy requirement, calibration can be achieved with slip gauges, micrometer or vernier calipers. In our instructions  

 

example the micrometer is suggested, which has similar accuracy to the dial gauge.

5.2.2. set the micrometer (25mm-50mm) to the target size "x", for example Ø45.75 and lock. In our example the tolerance required is H9 from 

 

Bs 4500:1969 (+.062/-0).

5.2.3. From instruction 5.1.2 and the tables on the back page; parts required are the anvil 44 plus spacer ring 2, from item 1 in fig.5.

 

Assemble as shown above in fig.1.

5.2.4. Maneouvre the gauge anvil and nib inside the measuring gap of the micrometer (fig.11). This activity could be eased by assistance or

 

carefully clamping the probe in a retort stand or similar.

5.2.5. observe the large dial pointer movement. It is essential that the axis of the gauge anvil and spring loaded nib are centralised in the    

 

micrometer measuring faces. It is essential to observe the total sweep of the dial pointer using the small dial. remember from the initial  

 

setting the pointer had rotated one full cycle. 

5.2.6. When satisfied with alignment, hold the position and rotate the dial bezel until the "0" [zero] aligns with the pointer. lock the bezel with   

 

the thumb screw. Your dial bore gauge has now been calibrated for use in a Ø45.25 to Ø46.25 bore ie +/-0.5mm of target.

5.2.7. The tolerance band can now be set using the two "markers" on the bezel. our example states the tolerance band to be +0.062 to -0. set  

 

one marker opposite and in line with "0" [zero] (fig.6) and the other +6.2 (fig.7) divisions apart. note! the larger the deflection the smaller  

 

the bore.

nib

Anvil

With a 2mm spacer

use the spacer rings to achieve "x" = 36, 37, 39, 40, 42 etc.. 

Insert the stylus for 1 full large pointer revolution

fig.5

fig.4

fig.3

fig.2

fig.1

small pointer on "1"

Be aware, nib projection is 2mm.

Aim for 0.5mm to 1.5mm as the

measurement zone.

spacer

n

ib total movement

Measurement zone

Pass

Pass

Fail

Fail

Fail

Fail

lower limit (+0.00mm)

upper limit (+0.062mm)

Example only; taken from 

Calibration text 5.2.

fig.6

fig.7

Tolerance

markers

  dBG509     Issue: 1 - 21/06/15

Original Language Version

 © Jack sealey limited

Содержание DBG509

Страница 1: ...R THE PURPOSE FOR WHICH IT IS INTENDED FAILURE TO DO SO MAY CAUSE DAMAGE AND OR PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE Thank you for purchasing a Sealey product Manufactured to a high standard this product will if used according to these instructions and properly maintained give you years of trouble free performance 2 introduction Precision mec...

Страница 2: ... anvil 44 plus spacer ring 2 from item 1 in fig 5 Assemble as shown above in fig 1 5 2 4 Maneouvre the gauge anvil and nib inside the measuring gap of the micrometer fig 11 This activity could be eased by assistance or carefully clamping the probe in a retort stand or similar 5 2 5 Observe the large dial pointer movement It is essential that the axis of the gauge anvil and spring loaded nib are ce...

Страница 3: ... anvil and the nib fig 11 rotate the barrel of the micrometer until the dial pointer aligns with the same 0 zero as the measured bore Take the reading from the micrometer and record 5 3 6 With the bore dial gauge now set the bore can now be measured for taper and ovality by reference back to the initial datum 0 zero reading and counting the 0 01mm divisions between datum and new Note the larger th...

Страница 4: ...e is 12 months from purchase date proof of which will be required for any claim 01284 757500 01284 703534 sales sealey co uk Sole UK Distributor Sealey Group Kempson Way Suffolk Business Park Bury St Edmunds Suffolk IP32 7AR www sealey co uk x Anvil size Spacer ring 34 5 35 5 35 No 35 5 36 5 35 1x1mm 36 5 37 5 35 1x2mm 37 5 38 5 38 No 38 5 39 5 38 1x1mm 39 5 40 5 38 1x2mm 40 5 41 5 41 No 41 5 42 5...

Отзывы: