7.0 Troubleshooting
Page 8
Do not attempt to service or adjust the instrument unless another person, capable of
providing first aid or resuscitation, is present.
If there are problems that cannot be resolved by the troubleshooting steps below please
contact technical support.
Technical Support
Tel: +1 (303) 665-1852 , Fax: +1 (303) 604-6088
[email protected], www.spectradynamics.com
AC Power LED does not turn on.
Disconnect AC and DC power cords. Remove the top cover. Check the AC power fuse and
power cord. If the fuse is blown replace with same type and rating.
DC Power LED does not turn on.
Disconnect AC and DC power cords. Remove the top cover. Check the DC power fuse
and power cord. If the fuse is blown replace with same type and rating.
Please contact SpectraDynamics if any of the fuses blows again or if the event that caused
the fuse to blow is not known.
Monitor LED is off.
Check to see if the RF signal provided to the instrument is greater than +7 dBm.
Check to see if a signal is present at all outputs of the signal distribution module.
The Monitor LED will remain off with a power lever under +7 dBm, but will still provide five
copies of the RF input signal. If the instrument is providing the copies of the input signal
you may continue using the DA-1G.
If the power level of the RF signal provided is greater than +7 dBm and the Monitor LED
remains off, the instrument will have to be returned for repair.