Warning: Incorrect IP address
Incorrect IP address entered.
Solution
Confirm the entered IP address, and enter
the correct IP address.
Error Message
Display
Description of Warning
E01: Test device error
Inserted test device is contaminated or
already used.
Solution
Discard the failed test device and retest with a
new test device.
E02: Insufficient sample volume
Insufficient sample is applied.
Solution
Discard the failed test device, and test again
by dropping sufficient volume of sample on
new test device.
E03: Expired test error
A test device was used after expiration date.
Solution
Perform the test again with new test device
that has not exceeded the expiration date.