Maintenance
The area surrounding the cable jacks at the top end of
the meter should be wiped with a clean cloth moistened
with alcohol to remove skin oils that will accumulate and
affect the accuracy at high resistances. The frequency
of cleaning will depend on usage; once a month would
be a good starting point.
Per ANSI/ESD S4.1 “Clean the electrodes with a
minimum 70% isopropanol-water solu tion.” Make sure
the conductive pads are dry prior to use.
See specific product test standard for test lab
specimen cleaning instructions. Per ANSI/ESD S4.1
Worksurfaces “The test specimens and electrodes
shall be cleaned twice with a minimum 70%
isopropanol-water solution using a clean, low-linting
cloth each time.” (Then conditioned for a minimum of 72
hours).
For compliance verification testing, do not clean
surfaces. However, if any measurements lie outside
acceptable range, then clean the material’s surface and
re-test.
Note: For worksurfaces, use Reztore™ Mat Cleaner or
other silicone-free ESD cleaner. Be sure the surface is
dry before testing.
The Surface Resistance Checker requires little
maintenance, and there are no user serviceable
parts. If your unit requires service beyond cleaning the
electrodes or replacing the batteries, please contact
SCS Customer Service
.
Specifications
Operating Temperature 50 to 95°F (10 to 35°C)
Environmental
Requirements
Indoor use only at altitudes
less than 6500 ft. (2 km)
Maximum relative humidity
of 80% up to 85°F (30°C)
decreasing linearly to 50% @
85°F (30°C)
Dimensions
5.2" L x 2.9" W x 1.1" D
(132 mm x 74 mm x 28 mm)
Weight
0.4 lbs. (0.2 kg)
Test Accuracy
±1/2 decade
Country of Origin
United States of America
Limited Warranty, Warranty Exclusions, Limit
of Liability and RMA Request Instructions
See the SCS Warranty -