Features and Components

A. Electrode Banana Jacks:

 Insert the test leads from 

the optional 5 pound electrodes here.

B. Electrode Toggle Switch:

 Toggle the switch to 

INTERNAL to measure with the parallel electrodes 

located on the back of the Surface Resistance Meter.  

Position the switch to EXTERNAL to measure with 

optional 5 pound electrodes (not included).

C. Resistance Measurement LEDs:

 The resistance is 

measured in ohms and read as 10

X

 ±1/2 decade where 

X is the range illuminated on the meter.

D. Parallel Electrodes:

 Ensure that the electrode toggle 

switch is set to INTERNAL when choosing to take a 

measurement with these electrodes.

E. Test Switch:

 Use this switch to make a surface 

resistance measurement.  Push and hold until one of 

the resistance measurement LEDs remains illuminated.

F. Low Battery LED:

 Replace the battery when this 

battery illuminates during use.  Do not use the Surface 

Resistance Meter when this LED illuminates.

G. Battery  Compartment:

 Remove the cover to allow 

access to the 9V battery compartment.

Operation

USING THE PARALLEL ELECTRODES
Measure Resistance Point-to-Point (Rtt) on the 

Surface

1.  Do not clean the surface. 
2.  Remove from the surface only those items that 

might interfere with the test. ESD sensitive devices 

shall also be removed. 

3.  Place the Surface Resistance Meter on the most 

commonly used area of the surface. The parallel 

electrodes should be 2” away from any edge and 

3” away from any grounded point. If the most used 

area is not obvious, use two points near the center 

of the surface. 

4.  Toggle the electrode switch located at the top of the 

Surface Reistance Meter to INTERNAL. 

5.  Push and hold the test switch until the measurement 

is displayed on one of the Resistance Measurement 

LEDs. 

6.  Perform additional measurements by placing the 

electrode on the most commonly used or worn area. 

 

NOTE: When taking multiple measurements, allow 

1-2 seconds between measurements. Pushing the 

test switch too quickly will result in an error.

If the measurement is outside acceptable limits, clean 

the surface with an ESD cleaner containing no insulative 

silicone such as Reztore® Surface & Mat Cleaner, 

and re-test to determine if the cause of failure is an 

insulative dirt layer or the ESD worksurface material.