Ground Man Plus Ground Monitor

Operation and Maintenance

 USER GUIDE TB-9026

Figure 1. Ground Man Plus Ground Monitor. 

Description

Overview

The SCS Ground Man Plus Ground Monitor can be 

used at work areas or locations to monitor the ground 

impedance of process tools including, but not limited to 

semiconductor, disk drive and flat panel environments. 

Ground impedance monitoring can be used for up to 

three points for metal ground within a tool. The Ground 

Man Plus Ground Monitor can also be used to monitor 

the resistance to ground and voltage level on a person. 

The unit incorporates green and red LEDs and audible 

alarms.
The Ground Man Plus Ground Monitor is available in two 

models:

SCS offers the following accessories for the Ground 

Man Plus Ground Monitor:

Packaging

1  Ground Man Plus Ground Monitor

1  Monitor Ground Cord (Green and Yellow)

1  Power Adapter, 12VDC, with interchangeable plugs    

    (North America, UK/Asia, Europe) (CTC334-WW only)

1  Certificate of Calibration

Ground Impedance Monitor 

Continuously monitors for ground impedance on three 

separate channels that are connected to three process 

tools. The alarm level is set the same for all channels.

Output Signal 

An output signal is available for connecting to the tool’s 

computer or PLC. The 5V signal is either logic 1 or 0 for 

indicating the monitored ground resistance/impedance 

conditions and for the wrist strap. 

Voltage on Person Wrist Strap Monitor 

The Ground Man Plus Ground Monitor has a feature 

that allows continuous monitoring of voltage on a person 

through a dual conductor wrist band assembly.

Performance

Impedance Alarm Level 

The alarm level can be preset from 1 to 20 ohms (from 

1 to 10 ohms in increments of 1 ohm and from 10 to 20 

ohms in increments of 2 ohms) by selecting through a 

combination of LED patterns indicated on the front on 

the unit. All three monitoring lines are set to the same 

selected impedance value. 

If the measured impedance on a line is lower than the 

setting, a green LED will remain illuminated for that 

line(s). If the impedance is higher than the preset level, a 

red LED and audible alarm are activated for that line(s). 

If the impedance level remains above the preset alarm 

level, the red LED and audible sound remains on.

The SCS Ground Man Plus Ground Monitor only 

monitors proper connection to ground. The monitored 

equipment must be properly ground by itself. A wire must 

be attached from the Ground Man Plus Ground Monitor 

(GND terminals) to approved ground points to monitor 

grounding of process tools correctly.

Output Signal Level 

The output connector provides a logic 1 (+5V) when all 

grounds are within limits and the unit is powered on. If 

any of the monitored grounds fail, a wrist strap fails, or 

power to the unit is lost, the output is logic 0 (0V).

Person Voltage and Resistance Monitor

The alarm level is monitored when a person is wearing 

a dual conductor wrist strap assembly and plugged into 

the front jack. There is a short hold on the body voltage 

 September 2015

Made in the

United States of America

Item

Description

2368

Dual Conductor Fabric Wrist Band, 

Adjustable

2360

Dual Conductor Ground Cord, Coiled, 5ft.

2370

Dual Conductor Ground Cord, Coiled, 10ft.

CTE701

Workstation Monitor Checker

CTA251

Replacement Jack PCB, Ground Man Plus

Item

Description

CTC334-WW

Ground Man Plus Ground Monitor, 

with World Wide Power Adapter

CTC334-T

Ground Man Plus Ground Monitor, 

without Power Adapter