SCS 770758 Скачать руководство пользователя страница 7

TB-9094

 Page 7 of 7

© 2019 DESCO INDUSTRIES INC

Employee Owned

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332

East: (919) 718-0000 | West: (909) 627-9634 • Website: 

StaticControl.com

Specifications

Dual Combination Tester

Input Voltage and 

Frequency

(External Adapter)

AC/DC Power Adapter
Power Input: 

100-240 VAC, 50/60 Hz
Power Output: 

12 VDC, 0.5 A
Cable Length: 

6 ft. (1.83 m)

Operating Temperature 70 to 85 °F (21 to 30 °C) for 1 

gigohm test limit
41 to 85 °F (5 to 30 °C) for all 

other test limits

Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity 

of 80 % up to 85 °F (30 °C) 

decreasing linearly to 50 % 

@ 85 °F (30 °C)
Maximum relative humidity of 

50 % at 1 gigohm setting

Dimensions 

(with mounting plate)

11.35" x 7.75" x 1.47"

(288 mm x 197 mm x 37 mm)

Weight

(with mounting plate)

1.2 lbs  (0.5 kg)

Test Accuracy

±20 % for 1 gigohm footwear 

test limit
±10 % for all other test limits

Test Switch Voltage

5 VDC @ open circuit

Wrist Strap and 

Footwear Test Voltage

30 VDC @ open circuit
Test current is limited by 

resistors and varies on 

the test range setting (100 

kilohms - 1 gigohm)

Relay Contact Rating

1 A @ 30 VDC max

Dual Foot Plate

Dimensions

13.25" x 15.25" x 1.125"

(337 mm x 387 mm x 29 mm)

Weight

5.5 lbs (2.5 kg)

Maintenance

To maintain optimum performance, cleaning should be 

performed on a regular basis. Use a minimum of 80 % 

Isopropyl alcohol to clean the foot plate and test switch. 

Other cleaners are susceptible to leaving residue on 

these surfaces.

Calibration

The Dual Combination Tester is calibrated to standards 

traceable to NIST. Frequency of recalibration should 

be based on the critical nature of those ESD sensitive 

items handled and the risk of failure for the ESD 

protective equipment and materials. In general, we 

recommend that calibration be performed annually.
The accuracy of the Dual Combination Tester is 

specified as:
±20 % for 1 gigohm footwear test limit

±10 % for all other test limits
A periodic check using a precision resistance box can 

be used to verify proper operation.
The SCS 

770751

 Limit Comparator is available for the 

convenient calibration of the Dual Combination Tester.
The Limit Comparator allows the customer to perform 

NIST traceable calibration on the Dual Combination 

Tester. The Limit Comparator can be used on the 

shop floor within a few minutes, virtually eliminating 

downtime, verifying that the Dual Combination Tester is 

operating within tolerances.
See 

TB-9070

 for more information.

Figure 12. SCS 770751 Limit Comparator

Limited Warranty, Warranty Exclusions, Limit 

of Liability and RMA Request Instructions

See the SCS Warranty - 

StaticControl.com/Limited-Warranty.aspx

Содержание 770758

Страница 1: ...Combination Tester 770751 Limit Comparator 770752 Power Adapter 100 240 VAC Input 12 VDC 0 5 A Output All Plugs Per ANSI ESD S20 20 section 7 3 A Compliance Verification Plan shall be established to e...

Страница 2: ...rd here to test your wrist strap F Dual Wire Wrist Strap Jack Insert your dual wire wrist cord here to test your wrist strap G Wrist Strap Status LEDs Displays the wrist strap test results H Relay Ter...

Страница 3: ...ON 10 Megohms 1 x 107 ON OFF 35 Megohms 3 5 x 107 default setting DIP switch 5 must be ON default setting for the wrist strap test to be active The wrist strap test will be disabled if DIP switch 5 i...

Страница 4: ...Secure the backplate using the two Phillips head screws and nuts 4 Plug the ground cord into the ground jack located on the bottom of the tester and connect the ring terminal to equipment ground This...

Страница 5: ...onding jack located on the face of the Dual Combination Tester Place one foot on each foot plate 3 To begin the test use your finger to bridge the test switch s inner and outer contacts The blue stand...

Страница 6: ...to ensure that they are being worn correctly and or need to be replaced NOTE Failures may be caused by dry skin or minimal sweat layer For wrist straps try using an approved dissipative hand lotion s...

Страница 7: ...bs 2 5 kg Maintenance To maintain optimum performance cleaning should be performed on a regular basis Use a minimum of 80 Isopropyl alcohol to clean the foot plate and test switch Other cleaners are s...