Balance
lO
w
Threshold
bRL
Balance Label
'"
•
Figure 16. Ionization balance screen
6. Push down the joystick momentarily to save all
settings.
"MEM"
will appear temporarily in the top
left comer while the settings are saved. The monitor
will then re-boot after the save is completed.
NOTE: The monitor will revert back to its last saved
settings if idled for 10 seconds while in the settings
menu. The joystick must be pushed down in order
to save all modifications to the settings.
Audible Alarm Settings
1. Toggle the joystick until the Sound channel is
displayed.
5nd
On
Figure 17. Sound On/Off screen
2. To change the audible alarm setting, push down
and hold the joystick until the setting below "Snd"
blinks.
5nd
'n
�
-un-
, '
Figure 18. Sound On/Off change screen
3. Toggle the joystick left or right to disable or enable
the audible alarm.
4. Push down the joystick momentarily to save all
settings. "MEM" will appear temporarily in the top
left comer while the settings are saved. The monitor
will then re-boot after the save is completed.
Smart Coefficient (SCF)
The Smart Coefficient (SCF) feature works in
conjunction with the static voltage field channel. This
feature is useful when an approximate voltage is
desired, especially when the distance is not exact. For
example, a known 1000 V static voltage source should
show approximately 1000 V on the EM Aware Monitor
when its antenna is 5 inches away from the static
voltage source. This 1000 V approximation will change
as the antenna is located further away from the static
voltage source. It may show a value close to 600 V
instead.
The Smart Coefficient function allows a correction factor
to compensate for the variable distance. To bring the
approximated voltage value back to 1000 V, the SCF
value will have to be set to 1.66. Alternatively, if the
antenna is moved closer, the monitor may show 1400 V.
In this case, an SCF value of 0.71 will bring the reading
back to 1000 V.
NOTE: All numbers referenced in the previous
paragraph are for illustration purposes only. They should
not be referenced for accuracy.
The Smart Coefficient has a default setting of 1.000,
and it can be adjusted from 0.001 to 1.999. When
adjusting the Smart Coefficient, it is important to have
a known static voltage source as a reference for setting
the SCF value. This voltage source may be a Pulse
Generator, a Charged Plate Monitor (1 kV typically), or a
Function Generator.
Smart Coefficient (SCF) Settings
1. Toggle the joystick until the Smart Coefficient
channel is displayed.
,nnn
I.LJ
LJ LJ
5[F
Figure 19.
Smart
Coefficient channel
2. To change the Smart Coefficient value, push down
and hold the joystick until the value blinks and
"REF" is displayed on the top-left corner. "REF"
indicates that the settings can be modified.
TB-9097
Page 16 of 19
© 2019 DESCO INDUSTR ES INC