Installation
21
GA01423_002_C0 - 10/2016 - © Leybold
3.3 Ambient Conditions
Danger of electric shock.
Install in an area which is not exposed to moisture such as rain or steam.
Moisture on electric power connections can cause bodily injury by short
circuit or electric shock.
Danger of explosion and fire.
Install in an area which is not exposed to explosives, flammable gas, or
other related hazards. Otherwise, the pump can cause an explosion or fire.
Operate at an ambient temperature of 5 °C - 40 °C. Operating outside of
this ambient temperature range can cause damage, fire or failure.
Ensure proper ventilation.
Install in a well-ventilated area. Poor ventilation can cause abnormal over-
heating, fire or failure since the scroll pump is air-cooled.
Necessary ventilation air:
SC 5 D
over 2 m
3
· min
-1
SC 15 D
over 4 m
3
· min
-1
SC 30 D
over 8 m
3
· min
-1
SC 60 D
over 12 m
3
· min
-1
Install in a proper location. Install in an area which is not exposed to dust or
corrosive gas. Failure to install in a proper location can result in damage or
failure.
Debris can clog silencers reducing the effectiveness of the purge gas.
Install in an area which is not exposed to sunshine. Direct sunshine can
cause high temperatures and failure.
DANGER
DANGER
Temperature
Avoid dust
Avoid direct sunlight