21
Troubleshooting / Technical specifications
5. Troubleshooting
If the wheelchair will not function while the batteries are fully charged, check the
following points before consulting your dealer:
Switch the controller off and then switch it on again. Check to see if the malfunction
has been solved.
Check if the free wheel switch was switched to Drive.
Check if the joystick was in the 0 position when the controller was switched on. In
other words, the joystick must not be moved when the controller is being switched ion
or off.
Malfunctions list
An extended malfunction list can be found in the appendix of the service manual (for
qualified specialists only). The service manual can be found on
www.scoutmobility.nl
6. Technical specifications
Supply Voltage:
24Vdc
Operating Voltage:
16Vdc to 35Vdc
Peak Voltage:
35Vdc
Reverse Battery Voltage:
40Vdc
PWM Frequency:
20kHz ± 0.5%
Brake Voltage:
12Vdc or 24Vdc
Brake Current:
100mA min.
1A max.
Charger Connector:
Use only Neutrik NC3MX
Batt. Charging Current:
12Arms max.
Actuator Current:
12A max.
Maximum Drive Current:
VR2 50 50A
VR2 60 60A
VR2 70 70A
VR2 90 90A
Moisture Resistance:
IPx4
Operating Temperature:
-25°C to +50°C
Storage Temperature:
-40°C to +65°C
EMC tested on sample wheelchair:
Susceptibility:
Tested at 30V/m to EN12184 (1999) and ANSI/
RESNA requirements
Emissions:
To EN55022 Class B
ESD:
IEC801 part 2