60
APPENDIX B: QUICK REFERENCE
087-0038 REV L
SCOTT SAFETY
Daily Bump Test
This section provides a brief amount of information on daily bump testing.
If installed, remove optional pump. Attach the Calibration Adapter and thread
Regulator to the calibration gas cylinder. Connect tubing. Apply gas. Device should
quickly respond to the target gases and begin alarming. Verify proper operation of
alarms. If device fails to respond (or is slow) or readings are not with in 10% of the
actual gas concentration, SPAN Calibration is required. See
.
Caution:
Do not operate the device with the Calibration Adapter attached.
Figure 42
Span Calibration - Calibration Adapter Attached
Daily Zero Calibration
This section provides a brief amount of information on daily zero calibration.
In clean atmosphere, press and hold the Right Operation button. Release when display
reads PLEASE WAIT. Device automatically perform ZERO calibration. Press Left
Operation button when display reads APPLY GAS to exit. See
Span Gas Calibration
This section provides a brief amount of information on span gas calibration.
The device’s calibration gas concentration as set in the Scott Gas Monitor (SGM)
software setting must match actual gas concentration used. Perform ZERO calibration
procedure connecting the Calibration Adapter with hose and regulator when APPLY
GAS displays. Once each sensor is calibrated, the device displays CAL PASSED then
REMOVE GAS. Remove Calibration Adapter and
wait
for the device’s main display
to return before turning off the device.
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...