CHAPTER 5: MAINTENANCE
087-0038 REV L
SCOTT SAFETY
Figure 28
Span Calibration - Calibration Adapter Attached
7
For single gas cylinders, wait for the display to read CAL PASSED then APPLY
GAS. Attach Tygon tubing to next cylinder and apply gas from the Regulator.
When final gas has been applied and passed calibration, REMOVE GAS displays.
Turn off gas cylinder and remove Calibration Adapter. The device automatically
returns to gas monitoring mode. See
.
8
For gas mixture cylinders, wait for the display to read CAL PASSED then
REMOVE GAS. Turn off gas cylinder and remove Calibration Adapter. The
device automatically returns to gas monitoring mode. See
.
If you power down the device while the LCD displays CAL PASSED and remove gas
the new calibration values are not saved. Ensure you wait until the LCD returns to the
gas monitoring mode.
Figure 29
Span Calibration - LCD Display Sequence
H
2
S ppm
LEL %
CO ppm
ZERO
H
2
S ppm
LEL %
CO ppm
O
2
%
H
2
S ppm
LEL %
CO ppm
CALIBRATE
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...