SYMPTON
POSSIBLE CAUSE
SUGGESTED CORRECTION
Wet ice
Ambinet temperature too high
Move unit to cooler location
High water level in the freezer
Lower to approx. 20 mm below
ice spout
Faulty compressor
Replace
Machine runs but makes no ice
Water not entering in the freezer
Air look in feed line to freezer.
Clogged feed line to freezer. Clean it
Gear stripped
Check and repair
Moisture in the system
Purge, replace drier and re-charge
Water leaks
Water seal leaking
Replace water seal
Water feed line to freezer leaking
Check and fasten hose clamp
Float valve not closing
Check and adjust float valve
setting screw
Excessive noise or chattering
Mineral or scale deposit on auger
Remove and manually polish auger
and inner freezer walls
and inner walls of freezer barrel
using emery paper
Low suction pressure
Add refrigerant to rise suction pressure
Water feed line to freezer clogged
Vent and clean it
Low water level into freezer
Adjust to approx. 20 mm below
ice spout
Gear motor noise
Worn rotor bearings
Check and replace
Shortage or poor lubricant in gear
Check for proper lubricant opening
case
gear case.
Top of gears must be covered
with lubricant
Gear case bearings and gear
Check and replace worn parts
racers worn out
Shortage of water
Strainer at water inlet fitting clogged
Remove strainer and clean
Float reservoir water nozzle
Remove float valve and clean
clogged-up
nozzle
SERVICE DIAGNOSIS