55
5)
Once you initiate the software you must select a template that is configured to include the
Aeroneb nebulizer; templates automatically included with the
flexiWare
software include
‘AN’ in their filename if they are configured for nebulizer use.
Before you proceed through calibration steps, the nebulizer is primed (run
continuously) per the settings in the template to create a seal through which air will not
escape. As such, be sure you have placed liquid in the nebulizer before proceeding
through this step.
6)
Place the nebulizer on the mount of the FX adapter and push down until it is flush with the
adapter base to ensure it is securely fastened.
7)
Proceed through the next steps of the start-up sequence (e.g. channel calibration).
Templates configured to include the Aeroneb often come pre-loaded with a script for an inhaled
dose response. If you opt to use this script (or a variation thereof), the triggering of the Aeroneb
is automated. You simply need to load it with the substance as needed during the script
execution.
To remove excess fluid from the Aeroneb reservoir, gently wick it up with a tissue. Take care
that no sharp tools come into contact with the nebulizer screen as this may damage the
Aeroneb.