3
1. Specifications
Maximum speed
4500rpm (300-4500rpm), increment: 100rpm
Maximum RCF
2490
×
g, increment: 100×g
Maximum capacity
10ml
×
12, 15ml
×
8
Timer
30 seconds -99 minutes-HOLD, continuous operation
Driving Motor
Brushless DC motor
Safety devices
Door interlock
、
over-speed detector
、
over-temperature detector
、
automatic
internal diagnosis
Power requirements
Single-phase,
110V-240V, 50Hz/60Hz, 3A
Dimensions
(
mm
)
(
L
)
255
×(
W
)
245
×(
H
)
140
Weight 6kgs
Additional features
Speed/RCF switch
、
Pulse operation
、
Processing display
、
Voice reminder
2. Declaration of Conformity
Construction in accordance with the following safety standards:
EN 61010-1
EN 61010-2-10
Construction in accordance with the following EMC standards:
EN 61326-1/ FCC Part 15 Subpart B/ IECS 001
Associated EU guidelines:
EMC-guidelines: 2004/108/EC
Instrument guidelines: 2006/95/EC
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be
required to correct the interference at his own expense.