• If they are not installed, then the standard interface components must be removed, and the NanoSpray
®
interface components installed. Refer to
®
®
Change to a Different Ion Source (OptiFlow
TM
Interface
Components)
Follow these steps to change from a NanoSpray
®
ion source to a Turbo V
TM
, IonDrive
TM
Turbo V, DuoSpray
TM
, or
PhotoSpray
®
ion source when the OptiFlow
TM
interface components are being used.
1. Remove the NanoSpray
®
ion source. Refer to
2. Remove the OptiFlow
TM
interface components. Refer to
3. Clean the standard curtain plate. Refer to the documentation that comes with the mass spectrometer.
Tip!
To easily clean the components before installing them on the mass spectrometer, remove the nano cell
curtain plate and nano cell heater assembly separately and then store them in the nano cell holder.
4. Install the standard curtain plate.
5. Install the ion source. Refer to the appropriate ion source
Operator Guide
.
Change to the NanoSpray
®
Ion Source (OptiFlow
TM
Interface
Components)
1. Remove the installed ion source. Refer to the
Operator Guide
for the ion source.
2. Install the OptiFlow
TM
interface components, by following these steps:
a. Clean the nano cell heater assembly. Refer to
Clean the Nano Cell Heater Assembly
b. Install the OptiFlow
TM
interface components. Refer to
.
3. Install the ion source. Refer to
.
Change to a Different Ion Source (NanoSpray
®
Interface
Components)
Follow these steps to change from a NanoSpray
®
ion source to a Turbo V
TM
, IonDrive
TM
Turbo V, DuoSpray
TM
, or
PhotoSpray
®
ion source when the NanoSpray
®
interface components are being used.
1. Remove the NanoSpray
®
ion source. Refer to
NanoSpray
®
Ion Source
Operator Guide
49 / 92
RUO-IDV-05-0811-F
Ion Source Maintenance