These factors can cause changes in ion source performance, indicating that maintenance is required.
Make sure that the installed ion source is fully sealed to the mass spectrometer with no evidence of gas leaks.
Regularly inspect the ion source and its fittings for leaks. Clean the ion source components regularly to keep the
ion source in good working condition.
CAUTION: Potential System Damage. Use only the recommended cleaning methods and
materials to avoid damaging the equipment.
Required Materials
• 1/4 inch open-ended wrench
• Flat-bladed screwdriver
• MS-grade methanol
• HPLC-grade deionized water
• Safety glasses
• Breathing mask and filter
• Powder-free gloves (nitrile or neoprene is recommended)
• Lab coat
Remove the Ion Source
Tip!
Before removing the source, note how the cables are routed, so that they can be routed the same way
when the source is installed.
The ion source can be removed quickly and easily, without tools. Always remove the ion source from the mass
spectrometer before you perform any maintenance on the ion source.
1. Stop any ongoing scans.
2. Shut down the sample stream.
3. Pull the X-Y-Z positioning unit as far back from the ion source interface as possible, until it is stopped by the
sleeve, to make sure that the supply of high-voltage power to the ion source head is disabled.
4. Deactivate the hardware profile.
5. Close the Analyst
®
/Analyst
®
TF software.
6. Let the ion source cool for 60 minutes.
NanoSpray
®
Ion Source
Operator Guide
45 / 92
RUO-IDV-05-0811-F
Ion Source Maintenance