![SCIEX NanoSpray Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/sciex/nanospray/nanospray_operators-manual_1217533046.webp)
Note:
For the NanoSpray
®
interface and OptiFlow
TM
interface wait for 60 minutes.
7. Turn off the illuminator by moving the switch to the position furthest from the LED.
8. Disconnect the 2-pin locking plug from the illuminator switch box.
9. Disconnect the camera video cable from the monitor video cable.
10. Disconnect the camera power cable from the 12 VDC power supply.
11. Disconnect the sample tubing from the ion source.
12. Turn the two source latches upward to release the ion source.
13. Pull the ion source gently away from the vacuum interface.
14. Put the ion source on a clean, secure surface.
Install the Ion Source
1. (For the TripleTOF
®
6600 mass spectrometer with SelexION
®
technology) If the OptiFlow
TM
interface is used,
then turn on the power to the SelexION
®
controller module.
The power switch is located on the back of the SelexION
®
controller module.
2. Align the ion source with the mass spectrometer. Make sure that the ion source latches are in their unlocked
positions (that is, the 12 o'clock position) and that they are aligned with the sockets on the mass spectrometer.
3. Push the ion source towards the vacuum interface and then turn the ion source latches towards the 6:00 o'clock
position until they stop. Do not force the latches after they become tight. Make sure that no gap is visible
between the ion source housing and the ion source interface.
CAUTION: Potential System Damage. Be careful when routing the camera and illuminator
cables. Make sure that they do not interfere with the movement of the X-Y-Z positioning
unit.
4. Connect the camera video cable to the monitor video cable. Refer to
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
46 / 92
Ion Source Maintenance