4.0 Maintenance
|9
4.0 Maintenance
WARNING: Do not tamper with any fixings other than
those indicated in the manual for mounting purposes.
There are no user serviceable parts within the MTP.
4.1 Troubleshooting
4.1.1 Movement not smooth
Check all the legs are set to the same height and all the screws are
tightened.
Remove micrometres from the MTP and check to see if the screw
housing has been damaged by over tightening.
Check to see if the micrometre rotates smoothly outside the MTP.
Move stage slowly back and forth without micrometres in place to check
for irregular movement.
4.1.2 No movement
Check shipping plates have been removed.
Check micrometres are properly in place and have been tightened.