User Manual SM6025
25
3.4
<LIST SWEEP DISP>
Up to 10 test frequencies, test levels or DC bias can be set on this page. Users can set the high and the
low limits for each list-sweep test point. Auto sweep test will be made on these test points. And their test
results will be compared with their limits.
Press down the menu key
[DISP]
and then the soft key
LIST SWEEP
to enter into the
<LIST SWEEP
DISP>
page, shown as below.
Test points will be automatically tested in a scanning mode. Meanwhile, comparison will be made
between test results and limit values. In the process of list sweep test, “
►
” denotes the current sweep
test point. The following control parameters can be set on
<LIST SWEEP DISP>
.
Sweep mode (MODE)
There are 2 zones on this page:
LIST SWEEP DISP
and
MODE
. List sweep points cannot be set on this
page but can be set on
<LIST SWEEP SETUP>
.
3.4.1 Sweep mode
The list sweep function of SM6025 can make automatic sweep test for up to 10 points’ test frequencies,
test levels or DC bias. Two sweep modes are available on SM6025 : SEQ and STEP. In SEQ mode, each
press of
[TRIGGER]
will direct SM6025 to automatically test all list sweep test points. In STEP mode,
each press of
[TRIGGER]
will direct SM6025 to test one list sweep point.
——————————————————————————————————————————————
NOTE
: When the trigger mode is set to INT, sweep test modes of SEQ and STEP will not be controlled by
[TRIGGER]
.
When the trigger mode is set to MAN, [TRIGGER] can be used to trigger the list sweep test.
——————————————————————————————————————————————
Operation steps for setting the list sweep mode
:
Set the sweep mode on the
<LIST SWEEP DISP>
page as
SEQ
or
STEP
.
1) On the
< LIST SWEEP DISP >
page, move the cursor to the MODE zone, the following soft keys will
be displayed:
SEQ
STEP
2) Press SEQ to set the sweep mode as sequential sweep test mode.