Operating Instructions
General Information
The logical front panel layout of
ensures rapid
familiarization with the various functions. However, even experienced operators
should not neglect to carefully read the following instructions, to avoid any
operational errors and to be fully acquainted with the instrument when later in use.
After unpacking the instrument, check for any mechanical damage or loose parts
inside. Should there be any transportation damage, inform the supplier
immediately and do not put the instrument into operation.
The case chassis and all measuring parts are connected to the protective earth
contact of the inlet. The mains plug shall only be inserted in a socket outlet
provided with a protective earth contact. The protective action must not be
negated by the use of an extension cord without a protective conductor.
When removing the metal case or replacing, the instrument must be completely
disconnected from the mains supply. If any measurement or calibration
procedures are unavoidable on the opened-up instrument, these must only be
carried out by qualified personnel acquainted with the danger involved.
SM5076 &
Safety
Warning!
Any interruption of the protective conductor inside or outside the
instrument or disconnection of the protective earth terminal is likely to
make the instrument dangerous. Intentional interruption is prohibited. The
mains/ line plug should be inserted before connections are made to
measuring circuits
.
SM
SM
SM
5076-1
Operating Conditions
5076 &
5076-1
The ambient temperature range during operation should be b 0 to +50 C
and should not exceed -40 C
or + 70 C during transport or storage. The
operational position is optional, however, the ventilation holes on the
must not be obstructed. Prior to calibration a preheat run of approx. 30
minutes is required
0
0
0
0