OPERATION GUIDELINES

UNIT SET UP

1. 

 Connect the power cord to the back of the unit 

and to an outlet.

2. 

 Turn the power switch found on the back of the 

unit to the On position.

3. 

 LED display shows 

10:00

. This is the most 

common timer setting.

If the timer needs to be adjusted,touch TIME+ key for 

quick timer settings and Time– key for fine adjustment 

between 1 to 30 minutes. Select desired time setting by 

pressing the TIME key.

Please consult your instrument cleaner instructions as 

well as your instrument directions for use for proper 

automated cleaning.
Prolonged cleaning time may result in:

1. 

Loosening of the screws if used.

2. 

 If items are plated and have existing cracks, the 

cracks may become more pronounced.

3. 

 Coatings with preexisting damages may 

experience more peeling.

DEGAS PROCEDURE

1. 

 Remove the cover and put it on the cover holder 

vertically. Add room temperature water to below 

the MAX but higher than the MIN fill lines.

2. 

 Add  OPT

IM

 28 Instrument Cleaning Solution per 

its Instructions for Use.

3. 

 Touch the Degas button and allow it to operate 

for 90 seconds.

 Please consult your Instrument cleaning Solution 

Instructions for Use to determine if degassing is 

required. OPT

IM

 28 does not require degassing. 

LOADING INSTRUMENTS

1. 

 Put items in cassettes or basket and lower into 

the cleaning tank.

2. 

 Confirm that the water level is between MIN and 

MAX and above the area to be cleaned.

3. 

 If not, add room temperature water to the tank. 

If the unit is turned on without water, ultrasonic 

energy will not be absorbed. Due to this 

unit’s high power, if on without water for over 

15 seconds damage to the unit may occur, 

severely reducing the life of the unit.

Содержание StatClean SC-P4862

Страница 1: ...ent control circuit for each ultrasonic transducer 40 higher cleaning efficiency not susceptible to interference more durable Multiple circuit protectors When overloaded or improperly used the protect...

Страница 2: ......

Страница 3: ...e air bubbles within the liquid which when in contact with an object dislodges the debris Using tap water is sufficient however it is recommended to use filtered water for best results Read the Manual...

Страница 4: ...y Precautions 1 Product Structure Accessories 3 Control Panel Operations 4 Operation Guidelines 5 Memory Operations 7 Common Cleaning Methods 8 Product Exploded View 10 Care and Maintenance 12 Specifi...

Страница 5: ...ved by the manufacturer Remove power cord by grasping its plug and not its cord Keep power cord away from obstructions cupboard doors chairs etc and hot surfaces to prevent cord damage Do not use devi...

Страница 6: ...ot use solution containing abrasive substances or strong corrosive chemical solution not recommended by the manufacturer or the supplier Operate unit on a dry and flat surface for operation When the u...

Страница 7: ...ol panel G Handlesa H Power switch I Power socket J Cover holder 1 Working time display 2 Multistage thermometer 3 Warning light 4 Degas status 5 Timer quick set key 6 Timer decreasing key 7 Normal cl...

Страница 8: ...Time Decreasing Key Touch key each time reduces the timer by 1 minute 8 Start Stop Key After powering on the LED display shows 10 00 which is the initial default time as shown and the unit is ready wi...

Страница 9: ...th preexisting damages may experience more peeling DEGAS PROCEDURE 1 Remove the cover and put it on the cover holder vertically Add room temperature water to below the MAX but higher than the MIN fill...

Страница 10: ...the cassettes or basket If additional cleaning is required reset the timer and repeat the process above OPERATION GUIDELINES DRAINING TANK At the end of the day the tank should be drained and cleaned...

Страница 11: ...er it will not remain if the power is off 4 If the Start Stop key is touched twice during a cycle the cycle and timer will stop It resumes when the Start Stop key is touched twice again NOTE To preven...

Страница 12: ...lease ensure you follow its instructions for use Suitable Locations Dental and medical offices and labs Cleaning Method 1 Place dirty items into basket and set basket inside the tank 2 Add water to th...

Страница 13: ...age valve to release the dirty water Close the valve afterwards 6 Clean the tank and add fresh water Wash the items for another 3 minutes to remove the residual debris and detergent D Special cleaning...

Страница 14: ...10 PRODUCT EXPLODED VIEW...

Страница 15: ...ndard part 22 Bracket ABS 23 Elbow PP 24 Valve handle PC 25 Bottom cover ABS 26 Main power PCB 94V0 27 Filter PCB 94V0 28 Cooling fan Standard part 29 Right handle ABS 30 Right panel ABS 31 Right cove...

Страница 16: ...ct sunlight for extended periods of time Keep the unit in a dry cool and ventilated area A thermal switch is attached to the bottom of the cleaning tank If the unit is turned on accidentally without w...

Страница 17: ...inum foil and cut 3 pieces 2cm 3 4 wide and 14cm 5 5 long Fold each piece to L shape along the length 6 Stick the 3 pieces evenly spaced from left to right across the center of the tank sticking the e...

Страница 18: ...ion StatClean Ultrasonic Cleaner Model SC P4862 Tank Capacity 6 0 L 1 6 gallon US Max 5 0 L 1 3 gallon US Min 3 35 L 0 9 gallon US Tank Size 32 3 x 20 5 x 9 9 cm 12 8 x 8 1 x 3 9 Longest Item Fits ins...

Отзывы: