Cleaning Method:

1. 

 Put the items to be cleaned in the recommended container. Add your instrument cleaning solution. 

We recommend OPT

IM

 28 Instrument Cleaner. Make sure you follow the instrument cleaning 

solution instructions for use.

2. 

 Place the container in the tank directly. Add water to a level between MIN and MAX at least half 

inch or 13mm above the bottom of the container but, not over the container.

3. 

 Turn the power switch on. Touch Time key to set the timer to 10 minutes. Consult your cleaner 

and/or instrument Directions for Use for proper time. Press Start/Stop key twice to start cleaning. 

Ultrasonic waves will penetrate the container and clean the items.

Selection of the container:

1. 

 Plastic containers – Soft and will absorb 30% - 40% of ultrasound waves, reducing the cleaning effect

2. 

Aluminum containers – Absorbs ~20% of ultrasonic energy.

3. 

Glass beakers – Absorb ~5-10% of ultrasonic energy

4. 

Stainless steel containers – Absorbs ~8% of ultrasonic energy.

8

COMMON CLEANING METHODS

A. Regular cleaning

Use filtered water for best results. You may add an 

instrument cleaning solution. We recommend  

OPT

IM

 28 Instrument Cleaner. Please ensure you follow 

its instructions for use.

Suitable Locations:

Dental and medical offices and labs.

Cleaning Method:

1. 

 Place dirty items into basket and set basket 

inside the tank. 

2. 

 Add water to the tank covering the area to be 

cleaned, keeping within the lines of MIN and 

MAX.

3. 

 Turn the power switch on. Press Time button 

to set the timer. Consult your cleaner and/or 

instrument instructions for use for proper time. 

Press Start/Stop button twice to start cleaning.

NOTE: 

 During cleaning, dirt will come off like a cloud and the water will gradually become murky. This is 

debris being removed from the instruments and indicates the instruments are being cleaned.

B. Indirect Cleaning

Put the items in a separate container.
Clean items using ultrasonic waves to penetrate the 
container.

Suitable Locations: 

Medical and dental instrument 

sterilization rooms. Different applications use different 

fluids to be used inside the container.

1. 

 Small instruments such as burs and files. Keep 

them separate from main instrument cleaning.

2. 

 Instruments that use a different solution such as 

cement remover.

3. 

 Instruments that use corrosive solution such as 

a disinfectant that’s corrosive. Avoid contacting 

the plastic housing with corrosive solutions.

4. 

 Multiple patients’ dentures or impressions. 

Disinfect the container after each patient.

Содержание StatClean SC-P4862

Страница 1: ...ent control circuit for each ultrasonic transducer 40 higher cleaning efficiency not susceptible to interference more durable Multiple circuit protectors When overloaded or improperly used the protect...

Страница 2: ......

Страница 3: ...e air bubbles within the liquid which when in contact with an object dislodges the debris Using tap water is sufficient however it is recommended to use filtered water for best results Read the Manual...

Страница 4: ...y Precautions 1 Product Structure Accessories 3 Control Panel Operations 4 Operation Guidelines 5 Memory Operations 7 Common Cleaning Methods 8 Product Exploded View 10 Care and Maintenance 12 Specifi...

Страница 5: ...ved by the manufacturer Remove power cord by grasping its plug and not its cord Keep power cord away from obstructions cupboard doors chairs etc and hot surfaces to prevent cord damage Do not use devi...

Страница 6: ...ot use solution containing abrasive substances or strong corrosive chemical solution not recommended by the manufacturer or the supplier Operate unit on a dry and flat surface for operation When the u...

Страница 7: ...ol panel G Handlesa H Power switch I Power socket J Cover holder 1 Working time display 2 Multistage thermometer 3 Warning light 4 Degas status 5 Timer quick set key 6 Timer decreasing key 7 Normal cl...

Страница 8: ...Time Decreasing Key Touch key each time reduces the timer by 1 minute 8 Start Stop Key After powering on the LED display shows 10 00 which is the initial default time as shown and the unit is ready wi...

Страница 9: ...th preexisting damages may experience more peeling DEGAS PROCEDURE 1 Remove the cover and put it on the cover holder vertically Add room temperature water to below the MAX but higher than the MIN fill...

Страница 10: ...the cassettes or basket If additional cleaning is required reset the timer and repeat the process above OPERATION GUIDELINES DRAINING TANK At the end of the day the tank should be drained and cleaned...

Страница 11: ...er it will not remain if the power is off 4 If the Start Stop key is touched twice during a cycle the cycle and timer will stop It resumes when the Start Stop key is touched twice again NOTE To preven...

Страница 12: ...lease ensure you follow its instructions for use Suitable Locations Dental and medical offices and labs Cleaning Method 1 Place dirty items into basket and set basket inside the tank 2 Add water to th...

Страница 13: ...age valve to release the dirty water Close the valve afterwards 6 Clean the tank and add fresh water Wash the items for another 3 minutes to remove the residual debris and detergent D Special cleaning...

Страница 14: ...10 PRODUCT EXPLODED VIEW...

Страница 15: ...ndard part 22 Bracket ABS 23 Elbow PP 24 Valve handle PC 25 Bottom cover ABS 26 Main power PCB 94V0 27 Filter PCB 94V0 28 Cooling fan Standard part 29 Right handle ABS 30 Right panel ABS 31 Right cove...

Страница 16: ...ct sunlight for extended periods of time Keep the unit in a dry cool and ventilated area A thermal switch is attached to the bottom of the cleaning tank If the unit is turned on accidentally without w...

Страница 17: ...inum foil and cut 3 pieces 2cm 3 4 wide and 14cm 5 5 long Fold each piece to L shape along the length 6 Stick the 3 pieces evenly spaced from left to right across the center of the tank sticking the e...

Страница 18: ...ion StatClean Ultrasonic Cleaner Model SC P4862 Tank Capacity 6 0 L 1 6 gallon US Max 5 0 L 1 3 gallon US Min 3 35 L 0 9 gallon US Tank Size 32 3 x 20 5 x 9 9 cm 12 8 x 8 1 x 3 9 Longest Item Fits ins...

Отзывы: