PEM Hydrogen generator quick instruction manual
Alarm message
Line
4
LCD
error
list:
Water empty error
Cause:
the water tank is empty
Action:
H2 generation stop,
alarm Light
blinks and buzzer sound every 10 second
Trouble shooting:
fill water tank with distillated water
Water over error
Cause:
the water tank is over full
Action:
H2 generation stop,
alarm Light
blinks and buzzer sound every 10 second
Trouble shooting:
empty water tank until alarm disappear
Pressure over error
Cause:
H2 pressure is more than 90 PSI
Action:
H2 generation stop,
alarm Light
blinks and buzzer sound every 10 second
Trouble shooting:
Turn instrument power off and call to company service section
Attention!!!
Change water and clean water tank every 1 month (30 days)