Venting requirements
Cabinet must have an air inlet vent at the bottom and outlet vent
no more than 200mm from the top
There must be a 100mm air space on
both sides of the under bench chiller
The front of the unit must have
at least 100mm of air space
The back of the unit needs
100mm of clearance for tubing.
The amount of air space around the unit apply for both passive and fan forced ventilation