Description of the programs
Progr. no. Name
Fixed parameters
Free parameters for own settings
1
Standards: Continuous stimulation with one frequency and one pulse width
1.0
Low frequency
2 Hz, 210 µs, 30 min
1-120 Hz, 50-300 µs, 10-90 min
1.1
High frequency
100 Hz, 120 µs, 30 min
1-120 Hz, 50-300 µs, 10-90 min
1.2
Gentle low frequency
2 Hz, 150 µs, 30 min
1-120 Hz, 50-300 µs, 10-90 min
1.3
Gentle high frequency
80 Hz, 70 µs, 30 min
1-120 Hz, 50-300 µs, 10-90 min
2
Han stimulation (according to Prof. Han): These programs run with 2 different phases (different frequen-
cies and pulse widths) which alternate every few seconds.
2.0
Han stimulation 1
Alternating 2 Hz, 210 µs, 3 s 100 Hz, 120 µs, 3 s, 30 min
2.1
Han stimulation 2
Alternating 2 Hz, 210 µs, 4 s 100 Hz, 120 µs, 2 s, 30 min
2.2
Han stimulation 2
Alternating 2 Hz, 210 µs, 2 s 100 Hz, 120 µs, 4 s, 30 min
3
Free Han stimulation: free parameter settings: These programs run with 2 different phases (different
frequencies and pulse widths) which alternate every few seconds.
3.0
Free Han stimulation 1
Alternating
2 Hz, 210 µs, 3 s 100 Hz,
120 µs, 3 s, 30 min
Alternating
F1: 1-120 Hz, 50-300 µs for 3 s
F2: 1-120 Hz, 50-300 µs for 3 s
10-90 min
3.1
Free Han stimulation 2
Alternating
2 Hz, 210 µs, 3 s 100 Hz,
120 µs, 3 s, 30 min
Alternating
F1: 1-120 Hz, 50-300 µs for 3 s
F2: 1-120 Hz, 50-300 µs for 3 s
10-90 min
3.2
Free Han stimulation 3
Alternating
2 Hz, 210 µs, 3 s 100 Hz,
120 µs, 3 s, 30 min
Alternating
F1: 1-120 Hz, 50-300 µs for 3 s
F2: 1-120 Hz, 50-300 µs for 3 s
10-90 min
4
Frequency modulation (FM) - The frequency is automatically modified within a specific range:
Min frequency - Max frequency - Min frequency
The pulse width is constantly adapted depending on the frequency.
4.0
Frequency modulation 1
60 Hz 120 Hz, 120 µs 100 µs in 60 s, 30 min
4.1
Frequency modulation 2
2 Hz 120 Hz, 210 µs 100 µs in 60 s, 30 min
4.2
Frequency modulation 3
2 Hz 15 Hz, 210 µs 200 µs in 60 s, 30 min
5
Pulse width modulation - The program uses a fixed frequency but the pulse width is automatically
modified within a specific range: Min pulse width - Max pulse width - Min pulse width
5.0
Pulse width modulation 1
2 Hz, 100 µs 210 µs in 120 s, 30 min
5.1
Pulse width modulation 2
15 Hz, 100 µs 210 µs in 120 s, 30 min
5.2
Pulse width modulation 3
80 Hz, 60 µs 150 µs in 120 s, 30 min
6
Burst - Burst programs use a high frequency stimulation in a low frequency rhythm: impulse packages of
100 Hz are applied for a fraction of time with pauses between the packages.
6.0
Burst 1
100 Hz (150 µs) impulses packages for 0.25 s,
then 0.25 s pause (2 Hz), 30 min
6.1
Burst 2
100 Hz (150 µs) impulse packages for 0.5 s,
then 0.5 s pause (1 Hz), 30 min
6.2
Burst 3
100 Hz (60 µs) impulse packages for 0.25 s,
then 0.25 s pause (2 Hz), 30 min