Technical data
Dimensions:
approx. 25 x 45 x 140mm
Swan neck:
approx.180 mm
Weight:
approx. 200g (including batteries)
Ex - marking:
see declaration of conformity
Energy supply:
2x batteries type Camelion Alcaline PLUS AA
(supplied)
Measurement range: 0 ppm to 2,5 Vol. %
Tolerance:
100ppm
-50ppm
+250ppm
with 20-50%rF
0,1Vol%
-250ppm
+0,5Vol%
with 20-50%rF
1,0Vol%
-0,2Vol%
+0,5Vol%
with 20-50%rF
2,2Vol%
-0,3Vol%
+0,2Vol%
with 20-50%rF
Resolution:
starting at 1 ppm
Operation time:
up to 10 hours.
The operation time will be significantly reduced by
the use of bad quality batteries or rechargeable
batteries.
Operating temp.:
0°C to +40°C
Stocking temp.:
-10°C to +50°C
Manufacturer: Schütz GmbH Messtechnik
- 21 –