220822
SCHUPP GmbH & Co KG
72280 Dornstetten-Aach | Glattal Straße 78 | T: +49 (0) 7443/243-0 | F: +49 (0) 7443/243-255 | [email protected] | www.schupp.shop
20
RUBIN
TROUBLESHOOTING
9.0 Assistance in case of malfunctions
Symptoms
Possible causes
Measures
Pads cannot be aligned
mechanically
If the couch has not been used
for a longer period of time, the
supporting gas pressure springs
can become stiff.
Gas strut defective.
In this case, the lever of the gas pressure spring
must be pulled completely upwards. With the
other hand, move the lying surface completely
up and down several times.
Exchange gas strut
No function of the couch drive
Motor is not connected to the
power supply
Fuse is defective
Cables are damaged
Check that the couch is properly connected to a
230V socket and that this socket is live.
Check the fuse in the house distribution box.
Check the power cable for damage.
Check supply line of hand or foot switch for
damage. Check whether the connecting cables
of the switching elements are correctly plugged
into the motor.
Engine noise
Contact the Schupp service department (see last
page).
No or limited function Hand or
foot switch
Wear
Pollution
Switching elements defective
Exchange hand / foot switch
Unplug the hand or foot switch from the motor.
Remove screws from cable cover,
Plug the of the new switching element back onto
the switch connector:
Replace cap and screws to secure.