Page 12
Operating
instructions
Cleaning the transducer head
The transducer head containing the sterile disposable nebulizing
chamber can be easily removed for cleaning and emptying by pulling it
upwards. Take care not to turn it.
The transducer head should be flushed out with clean water after every
use. Never immerse the transducer head completely in water!
For preventive disinfection of the transducer head, use commercially
available liquids such as antiseptic agents, Dodacarna S, Septolit or
Gigasept. The transducer head can also be flushed out with a 70%
alcohol solution.
The max. operating temperature of ceramic resonators (oscillating quartz
components) for the ultrasonic nebulizer is 85°C. Autoclaving damages
the transducer head.
Cleaning the nebulizer
The housing may be cleaned with a wipe disinfection.
Liquid agents such as Incidin Foam or Sekusept Aktiv may be used for
this. Make sure that no moisture enters the device.
NOTE
Do not clean the device while it is in use. To clean, turn the device off
and unplug it from the power line.