background image

 of 

1

13

Содержание LC SPRINT SPA

Страница 1: ...of 1 13...

Страница 2: ...CT DESCRIPTION 4 2 1 Components 4 2 2 Intended purpose 5 2 3 Intended use 5 2 4 Contraindications 5 2 5 Product variants and combinations 5 2 6 Material information 5 2 7 Operating life 6 3 INHALATION...

Страница 3: ...tionary measures which are to be observed to avoid damaging the product during use 1 4 Treatment of babies children and anyone who requires assistance Babies children and anyone who requires assistanc...

Страница 4: ...always use fresh water with reduced calcium content Make sure all components are dried properly whenever you have performed cleaning or disinfection Do not keep the product and accessories in a damp e...

Страница 5: ...r patients who are able to breathe by themselves and are conscious Only medication that has been approved for inhalation treatment must be used Take note of any restrictions in the instructions for us...

Страница 6: ...or seriously discoloured parts Please also follow the instructions for assembly given below Damaged components and or an incorrectly assembled nebuliser may impair functioning of the nebuliser and th...

Страница 7: ...Take care to ensure that the cap only moves in the direction allowed by the hinge Otherwise the nebuliser cap might snap off Insert the nebuliser in the holder on the compressor intended for this pur...

Страница 8: ...e treatment ensure that all parts are firmly connected to each other Several drops of medication will remain in the nebuliser after the treatment has ended Connect the connection tubing to the compres...

Страница 9: ...piratory valve out of the slot in the mouthpiece The valve must still be attached to the mouthpiece Rinse all components of the nebuliser under running tap water This must be done especially thoroughl...

Страница 10: ...parts The risk of infection is reduced when the parts are dried completely In boiling water Place all the individual parts in boiling water for at least 5 minutes Use a clean pot and fresh water with...

Страница 11: ...times when disinfecting Information about the amount of water required and about the disinfection process is included in the instructions for use of the devices in question disinfector and microwave C...

Страница 12: ...n and absorbent surface and let them dry completely Wrap the nebuliser in a clean lint free cloth e g a tea towel and keep it in a dry dust free environment 5 MISCELLANEOUS 5 1 Disposal All product co...

Страница 13: ...oduct information and in the event of defects or questions about usage please contact INQUA GmbH GERMANY inquair inqua de Manufacturer The product satisfies the basic requirements as set forth in Appe...

Отзывы: