Care and maintenance of the monitor
19
Care and maintenance of the monitor
Monitor test
Switch on the monitor once a week with the respiration sensor
plugged in but without applying the sensor to your child.
Ensure that the sensor lies completely undisturbed.
1. After the set time, the respiration alarm must occur (red LED
(2)) followed by the high-priority alarm tone.
2. Remove the respiration sensor. The fault display should now
light up (yellow LED (4)).
The tone will change to the medium-priority alarm tone.
=> Overview of alarm and fault indications Page 11
Cleaning Instructions for the device and Reusable Sensor.
Warning:
Do not autoclave or immerse the device or sensor in liquid.
Device:
Clean the surface of the device with a soft cloth moistened with
water or a mild soap solution. If disinfecting is necessary, wipe the
surfaces of the device with isopropyl alcohol (eg. Clinell
solution/wipes) Then wipe over with a soft water-moistened cloth.
Do not allow any liquid to enter any of the devices openings.
Sensor:
Disinfecting the resuable sensor – wipe/spray the sensor with
isopropyl alcohol following the disinfectant manufacturers guidleines
for relevant disinfection timescales. Then, wipe over with a soft
water-moistened cloth prior to use.