Overview of Alarm and Fault Indications
11
Overview of Alarm and Fault Indications
Significance of the LED colors
Green:
Activity display of the respiration function
Red:
Patient alarm: the permitted pause in respiration
has been exceeded
Yellow:
- Technical alarm: sensor error
- Battery alarm
Patient alarms are high-priority alarms.
They indicate a situation requiring immediate
attention to your child.
This includes the respiration alarm. It indicates that
the permitted pause in respiration has been
exceeded. The alarm is indicated by the red LED (2).
In addition, the sequence of tones for the patient
alarm is emitted repeatedly:
‘ Beep - Beep - Beep- short interval - Beep - Beep
– long interval ‘
=>Monitor behavior during patient alarm Page 17
Technical alarms are medium-priority alarms
These indicate a longer lasting fault or malfunction of
the monitoring channel, e.g. a defective or
disconnected respiration sensor.
Correct monitoring cannot be continued.
The alarm is indicated by the yellow LED (4).
In addition, the sequence of tones for the medium-
priority technical alarm is emitted repeatedly:
‘ Beep – Beep – Beep – long interval – Beep – Beep
–Beep – long interval ‘
=>
Monitor behavior during medium-priority technical
alarms Page 18