Remote Operation 

Basic use of the remote operation features are 
described in this section.  Cable lengths must be 
limited to a maximum length of 3m. 

Analog/Digital Controls 

The MC-LS may be operated remotely using a 
potentiometer and a switch connected to the 
appropriate pins of the IN/OUT port as shown in 
Figure 3. 
 
In the default configuration, the MC-LS must first be 
turned on using the front panel power switch.  
Connecting Pin 1 (Digital Input) to Pin 5 (Common) 
will then turn off the LED output. 

Serial Control (USB or RS-232) 

The MC-LS may be controlled remotely using 
simple text commands sent over the USB COM 
port or RS-232 serial port.  The RS-232 port must 
be configured for 9600 baud, 8 data bits, 1 stop bit, 
and no parity. 
 
All commands begin with an ampersand (&) and 
end with Carriage Return, ASCII code 0x0D 
(typically 

denoted as ‘\r’). 

 
Basic operating commands: 

 
 
 

Please refer to the MC-LS product page of the SCHOTT website for further information about remote operation. 

 

 

 

Figure 4: Outline Dimensions

 

 

 

9: +5V Ref

8: Fault Out

7: Not Used

6: Intensity

5: Common

4: Not Used

3: RS232 RX

2: RS232 TX

1: Digital In

 

 

Figure 2: IN/OUT Port Pinout 

 

 

 

Figure 3:  Analog/Digital Control, Typical Application 

 

Pin 

Description 

Digital input, 0-5V, 24V tolerant 

RS232 Transmit 

RS232 Receive 

Not connected 

Common return for all signals 

Intensity adjustment, 0-5V analog, 24V tolerant 

Not connected 

Fault output, open collector, 30V, 10mA 

+5V output, 10mA maximum 

 

10kΩ

Pin 9: +5V

Pin 5: Common

Pin 6: Intensity

Intensity 

Adjustment

Switch

Pin 1: Digital Input

On/Off 

Control

&L1 

Enable LED output 

&L0 

Disable LED output 

&I## 

Adjust LED intensity 
## must be a hexadecimal value between 
0 and FF (decimal 0 to 255) where FF is 
maximum brightness. 
Example: 

&I7F

 for 50% brightness 

 