13
Material Setup
Submenu
Settings Menu
Description
No. of the char-
acteristic mate-
rial curve 1 - 4
—
[character],
[numbers],
[special character]
to enter a name for the selected mate-
rial characteristic. The name can also
been adjusted by using the provided
software
Damping
—
[1] - [9];
[5]
Chapter 3.5.4
Alarms
—
[ON]
, [OFF]
To activate or deactivate the alarm
function for the material characteristic.
High limit
—
[0000] - [9999]
If the set limit value is exceeded, the
display reads
MAX-ALARM.
Low limit
—
[0000] - [9999]
If the value falls below the set limit
value, the display reads
MIN-ALARM.
Calibration
Chapter 3.6.2
Start
To perform a calibration, follow the in-
structions shown on the display.
Cal. Points
between [5 %] and
[90 %] of the
tension range
Set three calibration points for which a
calibration should be performed.
Weights
[cN],
[g]
Set the unit of the calibration weights
used.
3.4.1 Material menu
In the material menu you can make the settings for the selected material characteristics and
perform the calibration. To perform the calibration, the weights for the selected calibration
points must be available.
Factory settings are illustrated bold
3.4.2 Password
By setting a password, you can block the access to some menu parts for unauthorized
users. Therefor are 3 protection levels:
0 All menus are accessible (Factory setting)
1 The system setting menu is locked
2 The material, system setting, memory setting and cal. adjustment menus are locked
3 All menus are locked.
Apply the password protection
- Select the menu password (chapter 3.4)
- Insert a password and choose a protection level
- Exit the main menu, to apply the password protection
The factory setted password is „0000“
If you forgot the password, please contact HANS SCHMIDT & Co GmbH to request
the master password.