background image

5

2  Available Models 

2.1  Specifications

*  Corresponds to Shore A test procedure
** Pressure on test material surface when outer shell is in measuring position
  (colored line)

Model

Indenter

  mm Ø

Typical Applications

HP-2.5

Ball 2.5

Closely wound bobbins of synthetic fibers, etc.

HP-5

Ball 5

Loosely wound bobbins of synthetic fibers and closely 
wound natural fibers

HP-10

Ball 10

Very loosely wound bobbins of thick yarns, such as 
carpet yarns

HP-2.5-F

Flat 2.5

Same as Model HP-2.5 with a flat working face for 
measuring warp beams, or for bobbin diameters with 
over 400 mm

HP-5-F

Flat 5

Same as Model HP-5 with a flat working face for mea-
suring warp beams, or for bobbin diameters with over 
400 mm

HP-10-F

Flat 10

Same as Model HP-10 with a flat working face for 
measuring warp beams, or for bobbin diameters with 
over 400 mm

The principle used to measure the “firmness” or “hardness” of textile material is based on 
measuring the resistance force of the penetration of a ball sensor into a bobbin under a 
constant spring load. This principle is based on the Shore hardness test procedure.
The display range of the textile durometers covers a range from 0 to 100 units, where 0 
corresponds to the minimum and 100 to the maximum hardness value.
These units have no formal industry standards or engineering classification, and are not 
related to a specific Shore value or other known value. Since there are no international 
standards that can be used to compare against, we have developed a special test proce-
dure for verifying calibration (see Chapter 3.2).

General information on textile durometers

Model

HP-2.5, HP-5, HP-10

HP-2.5-F, HP-5-F, HP-10-F

Display Range

:

0 - 100 hardness units

0 - 100 hardness units

Depth of Indentation

:

0 - 2.5 mm

0 - 2.5 mm

Test Pressure

*:

approx. 12.5 N

approx. 12.5 N

Meas. Spring Force

:

0.55 - 8.065 N

0.55 - 8.065 N

Working Face Radius

:

55 mm

Working Face Dia.:

45 mm Ø

Scale Diameter

:

51 mm

51 mm

Weight

 net (gross):

approx. 300 g (appx. 500 g)

approx. 300 g (approx. 500 g)

Dimensions 

(LxWxH):

50 x 60 x 110 mm

50 x 60 x 110 mm

Содержание HP Series

Страница 1: ...HP Series Model HP 2 5 HP 5 HP 10 HP 2 5 F HP 5 F HP 10 F Edition HP 05 3 E Instruction Manual Valid as of 01 02 2010 Please keep the manual for future reference S C H M I D T c o n t r o l i n s t r u m e n t s ...

Страница 2: ...eclaration of Conformity 4 2 Available Models 5 2 1 Specifications 5 2 2 Delivery Includes 6 2 3 Unpacking 6 3 Operating Procedure 6 3 1 Notices Before Starting Measurement 6 3 2 Measuring with the Durometer 6 3 3 Verification of Calibration 7 3 3 1 Verify calibration of durometer models HP 2 5 HP 5 and HP 10 7 3 3 2 Verify calibration of durometer models HP 2 5 F HP 5 F and HP 10 F 8 4 Optional A...

Страница 3: ...e device Insufficient inspection of device components that are subject to wear Opening the device or improper repair work Disasters caused by the effects of foreign objects or by force majeure 1 1 Notices within the Operating Instructions The fundamental prerequisite for the safe handling of this device and its troublefree opera tion is the knowledge of the basic safety notices and safety instruct...

Страница 4: ... shall HANS SCHMIDT Co GmbH be held liable for damage resulting from misuse The intended use also includes Complying with all notices included in the Operating Instructions and observing all inspection and maintenance works 1 8 Copyright 1 9 Declaration of Conformity The copyright on these Operating Instructions remains with the company HANS SCHMIDT Co GmbH These Operating Instructions are intende...

Страница 5: ... on measuring the resistance force of the penetration of a ball sensor into a bobbin under a constant spring load This principle is based on the Shore hardness test procedure The display range of the textile durometers covers a range from 0 to 100 units where 0 corresponds to the minimum and 100 to the maximum hardness value These units have no formal industry standards or engineering classificati...

Страница 6: ...s that the same amount of pressure is applied from measurement to measurement and pre vents measuring errors Record the reading after approx 3 sec onds The hardness scale covers a range from 0 to 100 where 0 corre sponds to the minimum and 100 to the maximum hardness value Each test sample has to be tested in at least three different points Durometer Operating manual Specific Test Report 2 1 as pe...

Страница 7: ...aged by pressing the flat bot tomed models F models against a smooth level surface such as glass or marble When fully pressed down the scale pointer should make one full turn on the scale from zero to zero The scale pointer should return to the initial zero position when you release pressure If it does not return the instrument to the manufacturer for servicing On models with a concave base for me...

Страница 8: ... pointer should make one full turn on the scale from zero to zero If the result of verification is positive the durometer should show correct readings Models HP 2 5 F HP 5 F and HP 10 F Scale Glass plate Outer shell Color marking Holding screws fig 3 3 2 Do not loosen the holding screws of the movement under any circumstanc es Loosening these screws may alter the calibration of the instrument ...

Страница 9: ...s to zero the memory pointer remains in position marking the highest reading PEAK of the measurement 4 Optional Accessories Memory Pointer Option Code M Test Stand Model PSHP for serial testing Spring loaded lever for easy use Height adjustable from 0 to 200 mm Travel of operating lever 40 mm Dimensions LxWxH 162 x 100 x 320 mm Net weight approx 2 3 kg Base plate and prisms optionally available pl...

Страница 10: ...10 ...

Страница 11: ... the user s experience Assuming normal operating time and load as well as careful handling of the hardness tester we recommend a verification interval of 1 year 8 Correspondence Should you have any questions regarding the tension meter or Operating Instructions or their use please indicate above all the following details which are given on the ID plate 1 The tension meter model 2 The serial number...

Страница 12: ...one int 49 0 8638 9410 0 Fax int 49 0 8638 4825 int 49 0 8638 67898 e mail info hans schmidt com Internet http www hans schmidt com Tension Meter Force Gauge Torque Meter Tachometer Speed and Lengthmeter Electronic Lengthmeter Stroboscope Screen Printing Tension Meter Thickness Gauge Yarn Package Durometer and Shore A Durometer Sample Cutter Balance Moisture Meter Leak Tester Softometer More than ...

Отзывы: