5.1 WINDOWS Terminal Program
Commands for communication with a PC (polling)
The measured values can be transmitted to a personal computer. You can connect the
computer to the connector of the sensor by using the EK0640 special cable which is avail-
able as an accessory, respectively a USB cable.
The pin assignment of the RS-232 connector is described in Chapter 2.3.
Requirement:
A communication program, such as Terminal or HyperTerminal must be installed and con-
figured on the computer.
5 PC Communication
Setup commands:
Reading the ring buffer (only USB):
ASCII
Code
Function
Description
d
AVG value
current value
s
Fast Value
current value
z
zero
set zero
t
get information request the in-
formation of the
device (e. g. me-
auring range)
ASCII
Code
Function
g
integration 2 ms
h
integration 5 ms
i
integration 10 ms
j
integration 20 ms
k
integration 50 ms
l
integration 100 ms
m
integration 200 ms
n
integration 500 ms
Ring buffer response format:
@,Integrationtime in ms, comma separated values, CRLF
e. g. @,5,99,98,100,101,102,CRLF
The ring buffer contains max. 200 values
The instrument is easy to maintain. Depending on operating time and load, the instrument
should be checked according to the locally valid regulations and conditions.
6 Service and Maintenance