14
- Hang a weight that corresponds to e.g. 10 % of the tension range from the process
material.
- Start the measuring in the software
- Adjust the potentiometer, which you can reach through the ZERO hole in the housing,
with a screwdriver (with a point width of max. 1.9 mm) until the software displays the
corresponding value of 10 %,
Do not insert the screwdriver at an angle as this may damage
the potentiometer.
3.4.1 ZERO Adjustment
(Cont.)
Requirement
: ZERO adjustment carried out.
Thread the process material through the measuring and guide rollers, following the
material path symbol on the front of the sensor.
- Hang a weight that corresponds to e.g. 95 % of the tension range from the process
material.
- Start the measuring in the software
- Adjust the potentiometer, which you can reach through the GAIN hole in the housing,
with a screwdriver (with a point width of max. 1.9 mm) until the software displays the
corresponding value of 95 %,
3.4.2 GAIN Adjustment
Do not insert the screwdriver at an angle as this may damage
the potentiometer.