3.5.4 Using the alarm function
Requirements:
In the Material Setup menu, make sure that the MIN and MAX limit values have been set
for each material characteristic.
The limit value alarm has to be enabled in the main menu.
3.5.5 Cal. Adjustment
By performing a calibration adjustment, you can adjust a material characteristic
calibrated for a particular material to a different material or diameter without creating
a new material characteristic.
Requirements
:
- Prepare the measuring setup as shown in Fig. 3.5.5a.
- Select the desired material characteristic (chapter 3.4.1).
- Move the tension meter into the desired measuring position and carry out a zero
adjustment, if required.
Adjusting
:
- Select the „Cal. Adjustment“ menu item from the main menu (Fig. 3.5.5b)
- Insert the material to be measured (chapter 3.5.2) and push the thumbpiece into the
measuring position.
- Press the or key to perform the calibration adjustment until the value on the
display corresponds to the weight suspended to the material.
The adjustment can be performed in 1 % steps within the range from + 10 % to - 10 %.
- Press the
button to save the determined value.
Please note that this is a general value. So determine it separately for each mate-
rial characteristic and note it down for later reference. It is not possible to save it
per material characteristic.
To disable calibration adjustment, reset the value to 0 % in the „Cal. Adjustment“
menu.
fig. 3.5.5b
fig. 3.5.5a
Weight
Material to be
measured
Guide roller
Guide roller
Measuring
roller
The calibration adjustment has to be conducted with the weight unit, that is
adjusted in the menu „tension unit“ and must carried out with weights
corresponding to the choosen unit.