5. Information on use of special functions 

PULS 

μ

s / PULS deg

Flash duration. This function enables you to set the 

 ash duration. Using this function, you can 

in

 uence the brightness and focus of the object of observation. This adjustment can either be 

made in absolute form (microseconds) or in relative form (degrees).

DELAY ms

Adjustment of delay time between the internal trigger signal and the 

 ash (in milliseconds). This 

function enables you to set a 

 xed delay time between the internal trigger signal and the 

 ash.

Example: 

The position of observation can be adjusted extremely precisely without altering the 

 ash frequency. You can shift the observation position within a motion cycle. 

PHASE deg

Phase shift adjustment between the internal trigger signal and the 

 ash (in degrees, relative to 

the frequency). This function enables you to set a 

 xed angle between the internal trigger signal 

and the 

 ash.

Example: 

The position of observation can be adjusted extremely precisely without altering the 

 ash frequency. You can shift the observation position within a motion cycle. 

6. Changing the batteries

Using batteries/rechargeable batteries which are not completely charged reduces the operatio-

nal life until the next battery change. Never use or store the device with the battery compartment 

uncovered. 

  How to proceed:

•   Switch off device

•   Unscrew the screws on the back of the device

•   Remove batteries

•   Insert new, charged batteries

•   Dispose of old batteries in an appropriate way or recharge them (rechargeable batteries)

7. Scope of delivery

•  

S

troboscope, operating manual, certi

 cate, batteries, case 

  Additional scope of delivery for the version with trigger connection: 

•   Cable with plug for trigger signal

8. Accessories, optional

• Tripod adapter, tripod, calibration certi

 cate, belt pouch

4. Operation

NOTE! 

This chapter gives instructions for the operation of the version of the device without 

trigger connection. 

NOTE! 

Special functions and information on how to use the version of the device with trigger 

connection are described in chapter 9: „Additional operating information…“.

Position of buttons (Figure 1)

A = on/off. Hold down the button for approximately 3 seconds.

B = mode button. This button enables you to switch between different settings and modes in the 

order determined by the display (picture 7, numbers 1-10). 

NOTE! 

Some modes only have an effect on the functioning of the device when external trigge-

ring is employed.

C  = Increases the currently set value. Speeds up when this button is held down.* 

D  = Halves the currently set value. Speeds up when this button is held down.* 

E  = Doubles the currently set value. Speeds up when this button is held down.*

F  = Lowers the currently set value. Speeds up when this button is held down.*

The following functions are activated by simultaneously pressing the buttons shown below: 

B + F  = Reset to the default settings.

A + F  = Activate button lock / deactivate button lock. 

Display 

 elds (Figure 1)

In

 uencing the set signal. 

1    PULS 

μ

Flash duration (in microseconds).

2    PULS 

μ

Flash duration (in degrees).

3    DELAY ms   Delay time (in milliseconds) between the internal trigger signal and the 

 ash.

4    PHASE deg  Delay time between the internal trigger signal and the 

 ash (in degrees, 

       

 

relative to the frequency).

Display 

 7   Display. For unit – see numbers 8-10. 

Unit displayed 

 8   FPM  

Flashes per minute.

10   Hz   

Frequency of motion per second.

Operating information 

14 INT   

Device is generating 

 ash frequency.

15 LOBAT  

Battery running low.

NOTE!

 A parameter which has been set to differ from the default setting 

 ashes during operation. 

Содержание DSL Series

Страница 1: ...Stroboscope DSL Series Model DSL 300 DSL 300 T with trigger connection Operating Instructions S C H M I D T c o n t r o l i n s t r u m e n t s ...

Страница 2: ...ake technical changes without prior notice We have taken the utmost care to prevent errors Liability in the event of any errors shall be excluded Printed declaration of CE conformity on request A B C D F E 1 2 3 4 5 6 7 8 9 10 11 14 12 13 ...

Страница 3: ...pprox 11 h 6 000 FPM Disposable batteries approx 6 h 6 000 FPM Housing Material Aluminium ABS Heavy duty design Dimensions 191 x 82 x 60 mm Weight Approx 400g including batteries Ambient conditions Ambient temp 0 45 C Humidity Protection code IP65 Additional information for Stroboscope DSL 300 T with trigger input and output Trigger input Principle Optocoupler voltage free Low level 1 V Level 3 32...

Страница 4: ...ircumstances WARNING The device may not be used in areas where there is a danger of explosion WARNING Stroboscopic light can trigger epileptic ts in susceptible people ATTENTION Repairs on the device may only be carried out by the manufacturer or the supplier ATTENTION The electronic components of the device contain substances which are harmful to the environment They must be disposed of in compli...

Страница 5: ...tton is held down E Doubles the currently set value Speeds up when this button is held down F Lowers the currently set value Speeds up when this button is held down The following functions are activated by simultaneously pressing the buttons shown below B F Reset to the default settings A F Activate button lock deactivate button lock Display elds Figure 1 In uencing the set signal 1 PULS μs Flash ...

Страница 6: ...e to the frequency This function enables you to set a xed angle between the internal trigger signal and the ash Example The position of observation can be adjusted extremely precisely without altering the ash frequency You can shift the observation position within a motion cycle 6 Changing the batteries Using batteries rechargeable batteries which are not completely charged reduces the operatio na...

Страница 7: ...suitable for PNP and NPN signals A cable with plug corresponding to these input jacks is provided with the device Figure 2 Position of buttons Figure 1 The following functions are activated by simultaneously pressing the buttons shown below B D switch between internal and external trigger signal Display elds Figure 1 In uencing the input signal before the ash is generated 3 DELAY ms adjustment of ...

Страница 8: ...h too soon With the DELAY ms function you can set the value by which the output signal should be delayed PHASE deg Phase shift adjustment between the internal trigger signal and the ash in degrees relative to the frequency This function enables you to set a xed angle between the internal trigger signal and the ash Example The external trigger signal is generated before the required observation poi...

Отзывы: