page 30
cos105000-2016-00 en
Basic functions
Buttons / displays
Further information
Select “profile mode” with “+” or “-“
button.
Selected mode flashes.
Device must be in “mode selection”
(one of the mode LEDs flashes)
To get there, cancel all other
activities by pressing the “stop”-
button.
Confirm with “enter”.
Profile 1 is displayed.
Select profile with “+” or “-“.
Current profile is displayed in
“elevation” display.
Max. speed is displayed in “speed”
display.
Duration is displayed in “time”
display.
Confirm with “enter”.
The selected profile starts with the
first profile step after countdown.
The operation stops automatically
after the last step.
To scale the profiles, activate OP11
(User Options).
If scaling is performed, the max.
parameters are displayed (see
above).
Terminate the operation with “stop”.
For possibilities to interfere with an automatic program, see “interfere with an automatic program”.