SCHILLER BR-102 plus / BR-102 PWA
Patient Diary
Changing the batteries during a 48 hours recording
•
Preventive changing the batteries after 24 hours:
1. Press the blue button for 4 seconds. Confirm message
“Change battery”
again with the blue button.
2. Remove the batteries and replace with the fully charged batteries supplied. Observe correct polarity.
3. Press the green button to switch the unit on. The following message is displayed:
BP recording . . Next meas . . xx:xx
4. Make an entry in the patient diary.
•
Or when the audible beep is heard and one of the following messages are displayed:
1. Confirm the Change battery message with the blue button.
2. Remove the batteries and replace with the fully charged batteries supplied. Observe correct polarity.
3. Press the green button to switch the unit on. The following message is displayed:
BP recording . . Next meas . . xx:xx
4. Make an entry in the patient diary.
Do not dispose of the old batteries
- return to the surgery.
Cuff placement
The cuff should be placed on the left upper arm approximately 2cm above the elbow so that the forearm
can move. Position the MICRO symbol (if present) on the inside of the arm. The pressure tube points
towards the shoulder. Tape can be used to hold the tubing to the body if required. Place the device in the
pouch and secure with the pouch straps. Wear a t-shirt over the tubing to help secure the tubing in
position.
Make an entry in this diary (next page) if you experience any dizziness, palpitations, chest pain, etc. Also
make an entry when the batteries are changed, an extra measurement is taken, or for any other event.
BATTER Y L OW
Ch ang e
ba ttery
N O
YES
WAR NING
BATTER Y L OW
Ch ang e battery
imme dia te ly
OK
ERROR