BP recording
3.4
Long Term Recording
Page 38
BR-102 plus / BR-102 plus PWA
Art
. no.
:
2.
51
1076
Rev.
: g
3.4.4
Changing the Batteries During a 48 Hr Recording
For a 48 hour recording the batteries must be changed. An audible
alarm and visual indication is given when the batteries must be
changed.
Preventive changing the bat-
teries after 24 hours
1. Press the blue button for 4 seconds. Confirm message
“Change
battery”
again with the blue button.
2. Remove the batteries and replace with the fully charged batteries
supplied. Observe correct polarity.
3. Press the green button to switch the unit on. The following
message is displayed:
BP recording . Next measurement. . xx:xx
Changing batteries when
following audible and mes-
sages are displayed:
The batteries must be changed when an audible indication is given
and the message
Battery LOW - change battery
is displayed.
Proceed as follows:
1. Confirm
“Change battery”
with the blue button.
2. Remove the batteries and replace with the fully charged batteries
supplied. Observe correct polarity (
3. Press the green button to switch the unit on. The following
message is displayed:
BP recording . Next measurement. . xx:xx
An interrupted recording during battery change (unit is switched off),
is automatically continued when battery replacement occurs within 5
hours and the unit is switched on again.
Do you want
to change the
battery now?
NO
YES
CHANGE BATTERY
Battery low
CHANGE
Battery
NO
YES
WARNING
Battery low
Change battery
immediately
OK
ERROR