Firmware download
ARGUS VCM
16. After completion of downloading, turn off the monitor.
17. Disconnect Firmware downloading cable from the monitor and PC.
18. Press Power button and NIBP start/stop button simultaneously to enter
Service
mode
of the monitor.
19. Press Mode button to go to
DEFAULT RESET
.
20. Select DEFAULT RESET=YES to reset the monitor to factory defaults.
21. Turn off the monitor.
22. After a few seconds, turn the monitor on.
23. Perform the tests specified in
Performance Verification
section.
Note: Only Mediana provided Firmware download cable must be used. Otherwise,
Main BD may be damaged.
Note: Software versions of OEM boards are not field-upgradeable. (NIBP, SpO
2
or
Temperature module.) Replace the subsystem board with a higher (current)
version if necessary.
Note: If any problem during Firmware downloading, refer to Firmware Download in
Troubleshooting
section.
34
ARGUS VCM
Service Manual