Online Monitoring
The instantaneous SPL and 0.5s displayed SPL can be monitored. Firstly, set the interval
of logging in analyzer setup 2 to at least 100ms. And enter the measurement interface,
make sure the corresponding analyzer is open.
1/3 OCT: 12.5Hz, 16Hz, 20Hz, 25Hz, 31.5Hz, 40Hz, 50Hz, 63Hz, 80Hz, 100Hz, 125Hz,
160Hz,200Hz, 250Hz, 315Hz, 400Hz, 500Hz, 630Hz, 800Hz, 1kHz, 1.25kHz, 1.6kHz, 2kHz,
2.5kHz, 3.15kHz, 4kHz,5kHz, 6.3kHz, 8kHz, 10kHz, 12.5kHz, 16kHz, 20kHz, SPL(D), SPL(B),
SPL(A), SPL(C), SPL(Z)
ALL: LApeak, LCpeak, LZpeak, LA0.5, LAIp, LASp, LAFp, LC0.5, LCIp, LCSp, LCFp, LZ0.5
Note: The USB port communication method don't support for reading the basic infor-
mation
Note: 1. when the upper limit of y-axis is less than the max of the displayed parameters,
click the Y-axis to adjust the range of the Y-axis.
2. Don't swith the 0.5s and real-time directly, click the "Stop" firstly.
3. Double-click the center frequency in the graph, then the value of the SPL will be
displayed on the right side.
4.When download the data from the instrument, please don't operate the instrument
before finishing reading.
5. If you want to change the transmission method, like from RS232 to USB, please
reopen the software and reopen the instrument.