the washers, however, loosen all the screws partially
before loosening any one screw all the way.
3 • Remove the shield by pulling it slightly outwards at the bottom.
4 • To remove the lamp/reflector unit, gently squeeze it at both ends to
disengage the holes in the reflector from the four small pins in the
scanner lid assembly. Then pull it outwards bottom end first.
The old lamp/reflector unit can be discarded.
• NOTE
that the thermal paste on the small block behind
the lamp unit is normal and should
no
t be wiped off.
5 • To mount the new lamp/reflector unit, place the top fitting in the
holder in the scanner lid and,
gently squeezing both ends of the
reflector so the four holes can engange the four small pins in the
scanner lid, press the bottom fitting into the bottom holder.
CAUTION:
Make sure that all four pins have clicked into
the four holes.
CAUTION:
Do not touch the inner surface of the reflector!
6 • Place the outer shield in position and retighten the four Allen screws.
The end of the shield with the extensions is the one that
goes toward the front edge of the scanner lid.
7 • Plug in the scanner and turn it on.
8 • Check and, if necessary, realign the lamp assembly as described below.
Maintenance
ScanMate F6 / ScanMate F8 Plus • Operator’s Manual
5•3