5.5 Parts Counting

When the scale is showing weight, pressing the 

FUNC 

key will start the parts counting function. 

Before beginning, tare the weight of any container that will be used, leaving the empty container
on the scale. Place the number of samples on the scale. The number should match the options

for parts counting, 10, 20, 50, 100 or 200 pieces. 
Press the 

FUNC 

key to begin. The scale will show “SP 10" asking for a sample size of 10 parts.

Change the sample size by pressing the 

TARE/

 

key. the display will cycle through the

options: 10,20, 50, 100, 200 and back to 10. 
Press the 

ZERO/ENTER 

key when the number matches the number of parts used for the

sample. As more weight is added the display will show the number of parts (pcs). 
Press the 

key to display unit weight (g/pcs), Total weight (kg) or the count (pcs). Press the

FUNC 

key to return to normal weighing. 

5.6 Checkweighing

Check-weighing is a procedure to cause an alarm to sound when the weight on the scale meets
or exceeds values stored in memory. The memory holds values for a high limit and a low limit.

Either limit can be used or both can be used. 
Press the 

LIMIT 

key. The display will show the current High Limit with the left most digit flashing

and the HI symbol on to the left of the display.. 
To change the value shown use the 

%

/

 

and the 

LIMIT

/

 

to select the digit to change.

Then use the 

TARE

/

 

key to increment the flashing digit. When the desired value is shown

press the 

ZERO/ENTER 

key to accept the value. If you want to reset the value to zero press

the 

FUNC/C 

key to clear the value. 

After pressing the 

ZERO/ENTER 

key the display will then show the Low Limit, the LO symbol

will be on to the left side of the display. 
Enter the low limit in the same way the high limit was entered. 
After pressing the 

ZERO/ENTER 

key the scale will return to weighing with the Check-weighing

function enabled. 
When a weight is placed on the scale the arrows will show if the weight is above or below the

limits and the beeper will sound as described below. 
BOTH LIMITS SET 

The display will show OK and the beeper will sound when the weight is 

AHW/QHW digital

weighing scales user’s guide 

between the limits. 

LOW LIMIT SET, 
HIGH LIMIT is set to zero 

The display will show OK and the beeper will sound when the weight is less than the Low Limit.
Above the Low Limit the display will show HIGH and the beeper will be off. 

HIGH LIMIT SET, 
LOW LIMIT is set to zero 

The display will show LOW and the beeper will be off when the weight is less than the High
Limit. Above the High Limit the display will show OK and the beeper will be on. 

BOTH LIMITS SET. LOW IS SET GREATER THAN HIGH 
The beeper will never sound and the display will show LOW if the weight is less that the LOW

limit, and HIGH if the weight is greater than the Low Limit. 
NOTE: The weight must be greater than 20 scale divisions for the check weighing to operate. 
To disable the Check-Weighing function enter zero into both limits by pressing the 

FUNC/C 

key

when the current limits are shown then pressing the 

ZERO/ENTER 

key to store the zero values.

Содержание QHW+ 15

Страница 1: ...MANUAL QHW...

Страница 2: ...ntweighing sid 6 5 5 Parts counting sid 7 5 6 Checkweighing sid 7 5 7 Accumulated total sid 8 5 8 Manual accumulation sid 8 5 9 Automatic accumulation sid 8 6 Battery sid 9 7 RS232 sid 9 8 Parameters...

Страница 3: ...weight Entering a value using the keypad will store that value as the tare value A secondary function is increment to increase the active digit when in parameters or other functions Func Select the f...

Страница 4: ...4 1g 2g Units of Measure kg g Lb oz Interface RS 232 Optional Stabilisation time 2 seconds Operating temperature 0 C 40 C 32 F 104 F Power supply external 115 230 Vac 50 60Hz 10 watt Calibration Exter...

Страница 5: ...4 Displays The LCD display will show a value and a unit to the right of the digits In addition there are labels for TARE GROSS weight ZERO and for Low battery...

Страница 6: ...ss weight of the container plus all product that was removed The zero indicator will also be on because the platform is back to the same condition it was when the ZERO ENTER key was last pressed 5 3 W...

Страница 7: ...he ZERO ENTER key to accept the value If you want to reset the value to zero press the FUNC C key to clear the value After pressing the ZERO ENTER key the display will then show the Low Limit the LO s...

Страница 8: ...he scale is at zero The display will show the total number of items ACC xx and the total weight before returning to zero The totals will also be printed via the RS 232 interface To erase the memory pr...

Страница 9: ...ged As the battery is used it may fail to hold a full charge If the battery life becomes unacceptable then contact your distributor 7 RS232 The QHW Series of scales can be ordered with an optional RS...

Страница 10: ...select the sub menu press the TARE key then press the ZERO key to confirm the selected menu 9 Calibration 9 1 Normal calibration To start calibration turn the scale off and then turn it back on Press...

Страница 11: ...or Check the keys and connecter Err 6 A D value out of range Make sure platform empty and check the pan is installed proper Check the load cell connectors Err 9 Unstable Reading Check any air variatio...