8. Parameters

The scale has several parameters that can be set by the user. These allow the user to set the
scale to:

 

  Display the weight in other increments of weight to minimize the affects of vibration, wind or
other environmental conditions. 
  Control the back light on the display. It may be necessary to turn the backlight off to

maximize battery life. 
  Set the Accumulation to Automatic, manual or set the RS-232 interface to continuously print
the weight. 
  Set whether accumulation when print 
  Select another weighing unit(g) 

To set parameters, turn on the scale, press the 

FUNC 

and 

PRINT 

keys at the same time during

the self checking. 

The display will be show 

r 3 000

To select the menu, press 

LIMIT 

key and to select the sub menu press the 

TARE

/

 

key. then

press the 

ZERO 

key to confirm the selected menu. 

9. Calibration

9.1 Normal calibration

To start calibration turn the scale off and then turn it back on. Press the 

Tare 

and % keys

together during the initial counting from 9 to 0 on the display. 
The display will show "unLoAd". Remove all weight from the pan and then press the

ZERO/ENTER 

key to set the initial zero point. 

Then use 

%

Limi

t and 

Tare 

key to key in calibrate weight, press 

Zero 

key to sure, display

shows “load”, put this weight on platter, after stable indicator on, press 

Zero 

key to sure. 

After selfchecking again, AHW is ready for you. 

 

Calibrationweights 

QHW serien

Model #

QHW 3+

QHW 6+

QHW 15+

QHW 30+

Vikt 1

Zero

Zero

Zero

Zero

Vikt 2

1000g

2kg

5kg

10kg

Vikt 3

3000g

6kg

15kg

30kg

9.2 Internal counts

Press the 

ZERO 

and 

keys together during the initial counting from 9 to 0 on the display to

view the internal counts. 

Содержание QHW+ 15

Страница 1: ...MANUAL QHW...

Страница 2: ...ntweighing sid 6 5 5 Parts counting sid 7 5 6 Checkweighing sid 7 5 7 Accumulated total sid 8 5 8 Manual accumulation sid 8 5 9 Automatic accumulation sid 8 6 Battery sid 9 7 RS232 sid 9 8 Parameters...

Страница 3: ...weight Entering a value using the keypad will store that value as the tare value A secondary function is increment to increase the active digit when in parameters or other functions Func Select the f...

Страница 4: ...4 1g 2g Units of Measure kg g Lb oz Interface RS 232 Optional Stabilisation time 2 seconds Operating temperature 0 C 40 C 32 F 104 F Power supply external 115 230 Vac 50 60Hz 10 watt Calibration Exter...

Страница 5: ...4 Displays The LCD display will show a value and a unit to the right of the digits In addition there are labels for TARE GROSS weight ZERO and for Low battery...

Страница 6: ...ss weight of the container plus all product that was removed The zero indicator will also be on because the platform is back to the same condition it was when the ZERO ENTER key was last pressed 5 3 W...

Страница 7: ...he ZERO ENTER key to accept the value If you want to reset the value to zero press the FUNC C key to clear the value After pressing the ZERO ENTER key the display will then show the Low Limit the LO s...

Страница 8: ...he scale is at zero The display will show the total number of items ACC xx and the total weight before returning to zero The totals will also be printed via the RS 232 interface To erase the memory pr...

Страница 9: ...ged As the battery is used it may fail to hold a full charge If the battery life becomes unacceptable then contact your distributor 7 RS232 The QHW Series of scales can be ordered with an optional RS...

Страница 10: ...select the sub menu press the TARE key then press the ZERO key to confirm the selected menu 9 Calibration 9 1 Normal calibration To start calibration turn the scale off and then turn it back on Press...

Страница 11: ...or Check the keys and connecter Err 6 A D value out of range Make sure platform empty and check the pan is installed proper Check the load cell connectors Err 9 Unstable Reading Check any air variatio...

Отзывы: