Adjusting the alignment of the transmission lamp
1 • If the self-test shows that the transmission lamp is not properly aligned,
insert the special plastic-covered screwdriver (supplied with the scanner)
in the space between the scanner lid and the lamp shield to move the
lamp assembly
very slightly (ca. 1 mm) in the indicated direction:
•
front
end to the
right
= apply screwdriver at position
1
•
rear
end to the
right
= apply screwdriver at position
2
•
front
end to the
left
= apply screwdriver at position
3
•
rear
end to the
left
= apply screwdriver at position
4
CAUTION:
Use the screwdriver as a lever. Do not twist it.
2 • After making the adjustment, click on “Continue” to repeat the
alignment test.
3 • If correct alignment cannot be achieved after several tries, please
contact your ScanView dealer for assistance.
Maintenance
ScanMate F10 • Operator’s Manual
6•5
1
2
3
4
If adjustment is necessary, use the flat side of
the screwdriver as shown to move the lamp
assembly very slightly in the required direction.
Do not twist
the screwdriver.